Comparison of Arndt-endobronchial blocker plus laryngeal mask airway with left-sided double-lumen endobronchial tube in one-lung ventilation in thoracic surgery in the morbidly obese

نویسندگان

  • Z.J. Zhang
  • M.L. Zheng
  • Y. Nie
  • Z.Q. Niu
چکیده

This study aimed to evaluate the feasibility and performance of Arndt-endobronchial blocker (Arndt) combined with laryngeal mask airway (LMA) compared with left-sided double-lumen endobronchial tube (L-DLT) in morbidly obese patients in one-lung ventilation (OLV). In a prospective, randomized double-blind controlled clinical trial, 80 morbidly obese patients (ASA I-III, aged 20-70) undergoing general anesthesia for elective thoracic surgeries were randomly allocated into groups Arndt (n=40) and L-DLT (n=40). In group Arndt, a LMA™ Proseal was placed followed by an Arndt-endobronchial blocker. In group L-DLT, patients were intubated with a left-sided double-lumen endotracheal tube. Primary endpoints were the airway establishment, ease of insertion, oxygenation, lung collapse and surgical field exposure. Results showed similar ease of airway establishment and tube/device insertion between the two groups. Oxygen arterial pressure (PaO2) of patients in the Arndt group was significantly higher than L-DLT (154±46 vs 105±52 mmHg; P<0.05). Quality of lung collapse and surgical field exposure in the Arndt group was significantly better than L-DLT (effective rate 100 vs 90%; P<0.05). Duration of surgery and anesthesia were significantly shorter in the Arndt group (2.4±1.7 vs 3.1±1.8 and 2.8±1.9 vs 3.8±1.8 h, respectively; P<0.05). Incidence of hoarseness of voice and incidence and severity of throat pain at the post-anesthesia care unit and 12, 24, 48, and 72 h after surgery were significantly lower in the Arndt group (P<0.05). Findings suggested that Arndt-endobronchial blocker combined with LMA can serve as a promising alternative for morbidly obese patients in OLV in thoracic surgery.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

A novel combination of the Arndt endobronchial blocker and the laryngeal mask airway ProSeal™ provides one-lung ventilation for thoracic surgery

In this study, the feasibility and performance of the combination of the Arndt endobronchial blocker and the laryngeal mask airway (LMA) ProSeal™ in airway establishment, ventilation, oxygenation and lung isolation was evaluated. Fifty-five patients undergoing general anesthesia for elective thoracic surgeries were randomly allocated to group Arndt (n=26) or group double-lumen tube (DLT; n=29)....

متن کامل

One lung ventilation for children with Pierre Robin sequence using SupremeTM LMA and Arndt endobronchial blocker: a case report

One lung ventilation (OLV) is a technique routinely used in thoracic anesthesia to facilitate thoracic surgery. We describe a case of pediatric with Pierre Robin sequence using SupremeTM laryngeal mask airway (LMA) and Arndt endobronchial blocker to achieve OLV.

متن کامل

Airway injuries after one-lung ventilation: a comparison between double-lumen tube and endobronchial blocker: a randomized, prospective, controlled trial.

BACKGROUND Vocal cord injuries, postoperative hoarseness, and sore throat are common complications after general anesthesia. One-lung ventilation can be achieved via two techniques: double-lumen endotracheal tube or endobronchial blocker such as the Arndt blocker. The current study was designed to assess the impact of these techniques for one-lung ventilation on the incidence and severity of po...

متن کامل

Left mainstem bronchial rupture during one-lung ventilation with Robertshaw double lumen endobronchial tube -A case report-

Lung separation using a double-lumen endobronchial tube is necessary for video assisted thoracoscopy (VATs). Bronchial rupture after intubation with a double-lumen endobronchial tube has been rarely reported. We report a case of a 70-year-old man who had solitary pulmonary nodule in his right upper lobe. He was intubated with a left-sided Robertshaw double-lumen endobronchial tube. He underwent...

متن کامل

Measuring incisor-carina distance by fiberoptic bronchoscopy to guide the placement of a left-sided double-lumen endobronchial tube

The aim of this randomized observational study was to evaluate the effect of measuring the incisor-carina distance by fiberoptic bronchoscopy in guiding the placement of a left-sided double-lumen endobronchial tube. Fifty patients (American Society of Anesthesiologists class I-II) undergoing elective thoracic procedures requiring single-lung ventilation using a left-sided double-lumen endobronc...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره 51  شماره 

صفحات  -

تاریخ انتشار 2017