Heat and Moisture Exchangers and Humidification Efficacy in Pediatrics: Effects of Ventilator Settings and ETT Leakage
نویسندگان
چکیده
Heat and moisture exchangers (HMEs) have been increasingly used for humidification during mechanical ventilation in pediatric patients [1]. However, efficacy of HMEs in the pediatric population has not yet been fully evaluated and there is limited information in the literature regarding the effects of ventilator settings and endotracheal tube (ETT) leakage on the humidification performance of HMEs. Chikata et al. [2] tested ten pediatric HMEs in a model lung, measuring performance under different respiratory rates, tidal volumes TVs (achieved with different pressure control levels), and leakage conditions. They found that eight of the ten HMEs maintained absolute humidity (AH) at more than 30mg/L, the minimum level recommended by the American Association for Respiratory Care (AARC). With a small leak, AH decreased below 30mg/L (26.6 to 29.5mg/L), decreasing further (19.7 to 27.3mg/L) with a larger leak. Pressure control (tidal volume) level did not statistically significantly affect AH values, although there was a trend to decreased AH at higher tidal volumes. In three of the HMEs, increased respiratory rate resulted in a significantly higher AH. In six of the other seven HMEs, there was a nonstatistically significant trend to higher AH with increased respiratory rate, with one HME showing a nonsignificant trend to lower AH. We think this is a well-designed and conducted study and provides valuable information regarding the use of HMEs in pediatric patients under different conditions. The effect of ETT leakage on AH is nicely demonstrated and is consistent with previously published data on mechanically ventilated adults. However, we had some questions regarding the study’s other findings and their possible explanation as follows
منابع مشابه
Humidification Performance of Heat and Moisture Exchangers for Pediatric Use
Background. While heat and moisture exchangers (HMEs) have been increasingly used for humidification during mechanical ventilation, the efficacy of pediatric HMEs has not yet been fully evaluated. Methods. We tested ten pediatric HMEs when mechanically ventilating a model lung at respiratory rates of 20 and 30 breaths/min and pressure control of 10, 15, and 20 cmH(2)O. The expiratory gas passed...
متن کاملEvaluating Humidity Recovery Efficiency of Currently Available Heat and Moisture Exchangers: A Respiratory System Model Study
OBJECTIVES To evaluate and compare the efficiency of humidification in available heat and moisture exchanger models under conditions of varying tidal volume, respiratory rate, and flow rate. INTRODUCTION Inspired gases are routinely preconditioned by heat and moisture exchangers to provide a heat and water content similar to that provided normally by the nose and upper airways. The absolute h...
متن کاملInfluence of ambient temperature and minute ventilation on passive and active heat and moisture exchangers.
OBJECTIVE During invasive mechanical ventilation, inspired gases must be humidified. We previously showed that high ambient temperature greatly impaired the hygrometric performance of heated wire-heated humidifiers. The aim of this bench and clinical study was to assess the humidification performance of passive and active heat and moisture exchangers (HMEs) and the impact of ambient temperature...
متن کاملEffect of heat and moisture exchanger (HME) positioning on inspiratory gas humidification
BACKGROUND In mechanically ventilated patients, we investigated how positioning the heat and moisture exchanger (HME) at different places on the ventilator circuit affected inspiratory gas humidification. METHODS Absolute humidity (AH) and temperature (TEMP) at the proximal end of endotracheal tube (ETT) were measured in ten mechanically ventilated patients. The HME was connected either direc...
متن کاملHumidification on Ventilated Patients: Heated Humidifications or Heat and Moisture Exchangers?
The normal physiology of conditioning of inspired gases is altered when the patient requires an artificial airway access and an invasive mechanical ventilation (IMV). The endotracheal tube (ETT) removes the natural mechanisms of filtration, humidification and warming of inspired air. Despite the noninvasive ventilation (NIMV) in the upper airways, humidification of inspired gas may not be optim...