Comparison of Doppler echocardiographic methods with heart catheterisation in assessing aortic valve area in 100 patients with aortic stenosis.
نویسندگان
چکیده
OBJECTIVE To examine the practicability and accuracy of Doppler echocardiographic methods in determining aortic valve area. METHODS Aortic valve areas determined by three methods using Doppler echocardiography (applying the continuity equation and the modified Gorlin formula using data from Doppler echocardiography and right heart catheterisation) were compared with values obtained by heart catheterisation. PATIENTS 100 consecutive patients with aortic stenosis aged between 34 and 83 years (mean (SD) 66 (10)). RESULTS Differences in individual patients' measurements of aortic valve area by the three Doppler techniques varied by up to 0.56 cm2 compared with values obtained by heart catheterisation. On average, values obtained from Doppler echocardiographic methods lay up to 51% below and 78% above those obtained by heart catheterisation. CONCLUSIONS All three Doppler echocardiographic methods were practicable in routine clinical practice for patients of all ages, but they were of limited accuracy when compared with the aortic valve areas found invasively using the invasive Gorlin equation. However, these deviations may not always be due to inadequacies of the Doppler methods: they could also be caused by limitations in the Gorlin formula. Doppler methods can be repeated if required, they allow examination of the morphology of the valve, and they subject the patient to considerably fewer risks than the invasive procedure. An adequate strategy in determining the severity of aortic valve stenosis would be to calculate the valve area by Doppler echocardiography as well as considering the valvar aortic pressure gradient. The valve area alone should not be relied on exclusively, as has been the increasing practice in the past few years.
منابع مشابه
Doppler echocardiographic findings in tissue engineered aortic valve in a sheep model
Background: Heart valve diseases are considered a common disease in human and animals, and valve replacement is an option for treatment of valvular diseases. Objectives: In this study efficacy of a tissue engineered valve in thoracic aorta was evaluated with transthoracic echocardiography. Methods: This study was undertaken on 6 male sheep. Echocardiography was performed on all sheep 24 hours b...
متن کاملLimitations in assessing the severity of aortic stenosis by Doppler gradients.
Continuous wave Doppler echocardiography was performed before cardiac catheterisation in 69 consecutive patients with suspected aortic stenosis. Agreement between the maximum and the mean Doppler gradients and catheterisation gradients was good. Doppler echocardiography, however, systematically underestimated the maximum and mean gradients, particularly in the high range. Stepwise regression an...
متن کاملA new non-invasive estimation of the stenotic aortic valve area by pulsed Doppler mapping.
A new pulsed Doppler mapping technique has been used to measure the severity of aortic valve stenosis. The Doppler examination was performed at the site of the aortic orifice in the parasternal short axis echocardiographic view and the method was based on the detection of the area of systolic flow through the stenotic orifice. This area was derived by planimetry and the measurements obtained by...
متن کاملPlanimetry of aortic valve area using multiplane transoesophageal echocardiography is not a reliable method for assessing severity of aortic stenosis.
OBJECTIVE To assess the reliability of aortic valve area planimetry by multiplane transoesophageal echocardiography (TOE) in aortic stenosis. DESIGN Study of the diagnostic value of aortic valve area planimetry using multiplane TOE, compared with catheterisation and the continuity equation, both being considered as criterion standards. SETTING University hospital. PATIENTS 49 consecutive ...
متن کاملمقایسه نتایج درمانی بالون والوپلاستی و والوتومی جراحی در بیماران مبتلا به تنگی مادرزادی دریچه آئورت در پیگیری 6 ساله در بخش کودکان بیمارستان قلب شهید رجایی تهران
Background: Balloon valvuloplasty is a relatively new technique for relief of obstruction of aortic valve. The primary results of this procedure are similar to surgical valvotomy with fewer complications and mortality. Materials and methods: One hundred fifty-five patients with congenital aortic stenosis were enrolled in this study over a period of 6 years (1991-1997). Participants were all un...