Magnetic Sector ICPMS with Desolvating Micronebulization: Interference-Free Subpicogram Determination of Rare Earth Elements in Natural Samples.
نویسندگان
چکیده
A new method has been developed combining desolvating micronebulization with magnetic sector inductively coupled plasma mass spectrometry (ICPMS) for the analysis of all 14 stable rare earth elements (REEs) in small samples of marine particulate matter. Application is demonstrated for REEs in suspended particles from a deep ocean hydrothermal vent plume and a geological reference material. A 100-fold reduction in oxide formation, relative to standard nebulizer-spray chamber sample introduction, makes oxide interference correction negligible, even for samples that are very enriched in Ba and light REEs. Enriched isotopes for one light and one heavy REE ((145)Nd and (171)Yb) are used as both isotope dilution and internal standards, providing determination of all the REEs in one analysis. This standardization scheme eliminates the need for multimass drift correction used previously to achieve acceptable accuracy with external standardization techniques. Instead, the method exploits capabilities for accurate and precise determination of isotope ratios, a principal strength of ICPMS, and the mass-independent sensitivity of electric field scans on our double-focusing instrument. We demonstrate overall precision of ≤2% (1σ) and accuracy better than 6% for all the REEs (except Er = 8.7%), based on comparison to recommended values for USGS certified reference material BHVO-1 (basalt). This performance is similar to that obtained by full isotope dilution mass spectrometric techniques, but the new method is far simpler, requires 5 min sample(-)(1), and avoids interferences introduced by complex mixtures of enriched isotopes. Sensitivity of (1.2-1.4) × 10(6) counts s(-)(1) ppb(-)(1) and background intensities of 2-60 counts s(-)(1) provide excellent detection limits of 1-40 ppq, a 100-fold improvement on established ICPMS methods. The low sample introduction rate (100 μL min(-)(1)) allows unprecedented absolute detection limits of 1-20 fg.
منابع مشابه
Direct determination of 10 trace metals in 50 mL samples of coastal seawater using desolvating micronebulization sector field ICP-MS†
Understanding the trace metal marine geochemistry of temporally variable coastal systems requires intensive sampling programs with attendant analytical burdens. Most established techniques for multi-element trace metal determinations are slow, require a skilled chemist, and are not easily automated. Advances in sample introduction systems and ICP-MS instrumentation now provide marine chemists w...
متن کاملDetermination of Rare Earth Elements in Products of Chadormalu Iron Ore Concentrator Plant (Iran) from Beneficiation Point of View
Different samples have been prepared from different products in Chadormalu iron ore concentrator plant: Low intensity magnetite separators concentrate (magnetite concentrate), reverse flotation tail (final hematite concentrate), flotation concentrate (apatite concentrate), final tail (L.I.M.S. tail + reverse flotation concentrate+ apatite flotation tail). The samples were used for rare eart...
متن کاملTrace Element Microanalysis in Iron Meteorites by Laser Ablation ICPMS.
A laser ablation microanalysis system has been developed that can analyze trace elements with a sensitivity in the ppb range, using a CETAC LSX-200 laser ablation system with a Finnigan Element. This capability has been applied to a set of iron meteorites to demonstrate the laser microprobe's analytical capability for the determination of platinum group elements (PGEs) with a spatial resolution...
متن کاملTrace analyses of arsenic in drinking water by inductively coupled plasma mass spectrometry: high resolution versus hydride generation.
A magnetic sector inductively coupled plasma mass spectrometer (ICPMS) was applied to the determination of arsenic in drinking water samples using standard liquid sample introduction in the high-resolution mode (M/delta M = 7800) and hydride generation in the low-resolution mode (M/delta M = 300). Although high mass resolution ICPMS allowed the spectral separation of the argon chloride interfer...
متن کاملDetection of Main Rock Type for Rare Earth Elements (REEs) Mineralization Using Staged Factor and Fractal Analysis in Gazestan Iron-Apatite Deposit, Central Iran
Gazestan magnetite-apatite deposit is located in Central Iran and Bafq region, which has been occurred in form of veins, veinlets, and small apatite lenses as well as magnetite in metasomatic rock types such as green chlorite-actinolite rock units. These rocks are situated in the carbonate-volcanic complex of Upper Precambrian-Lower Cambrian Rizo formation. In this study, staged factor analysis...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
- Analytical chemistry
دوره 70 21 شماره
صفحات -
تاریخ انتشار 1998