Accurate multislice gradient echo T(1) measurement in the presence of non-ideal RF pulse shape and RF field nonuniformity.

نویسندگان

  • G J Parker
  • G J Barker
  • P S Tofts
چکیده

A methodology is presented for estimating T(1) using a two-point technique with a standard multislice gradient echo sequence. The method explicitly corrects for the shape of the RF pulse and for spatial variations in transmitted RF field intensity. Including these factors in the calculation of T(1) gives a substantial improvement in accuracy and precision of two-point gradient echo T(1) measurements in the presence of RF nonuniformity and non-ideal pulse profiles. The mean accuracy of the technique was found to be 1.4%. The mean precision was found to be 3.0%. The mean variation along the axis of a head coil was found to be 2.3%.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Assessment of the Characteristics of MRI Coils in Terms of RF Non-Homogeneity Using Routine Spin Echo Sequences

Introduction: One of the major causes of image non-uniformity in MRI is due to the existence of  non-homogeneity in RF receive and transmit. This can be the most effective source of error in  quantitative  studies  in  MRI  imaging.  Part  of  this  non-homogeneity  demonstrates  the  characteristics of RF coil and part of it is due to the interaction of RF field with the material being  imaged...

متن کامل

اندازه گیری غیریکنواختی امواج رادیوئی در ام آر آی

Introduction: Non-uniformity is one of the most important parameters affecting MRI images which can lead to harmful effects in the diagnosis and analysis of qualitative and quantitative methods. The present study introduced a method for measuring RF non-homogeneity in MRI systems. Methods and Materials: To verify the uniformity of B0 and B1 fields, a cylindrical phantom with a diameter of 24 c...

متن کامل

Simultaneous multislice imaging with slice-multiplexed RF pulses.

A method for simultaneous multislice imaging is presented that uses a multislice RF pulse that imparts a different linear phase profile to each slice. During readout, slices are unaliased by using extra slice-select gradient lobes, which rephase and dephase individual slices one at a time. Compared to other simultaneous slice methods, this method avoids distortion by slice-select gradients, and...

متن کامل

Wideband Parallel Imaging

Human lower limb images were successfully acquired with an acceleration rate of up to 8x by using wideband parallel imaging. The non-uniform multislice profile problem of this technique was successfully compensated to enhance the image quality of the outer slices. Also the signal-to-noise ratio (SNR) of this ultra fast imaging technique, wideband parallel imaging, was verified to be similar wit...

متن کامل

Sources of intensity nonuniformity in spin echo images at 1.5 T.

This manuscript describes a number of sources of nonuniformity for spin echo images at 1.5 T. Both coil tuning and crosstalk can have significant effects on image nonuniformity. For short repetition times, nonuniformity increases with decreasing TR, possibly due to gradient eddy currents. In sections of RF coils with poor RF uniformity, image nonuniformity varies with both echo time and the num...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:
  • Magnetic resonance in medicine

دوره 45 5  شماره 

صفحات  -

تاریخ انتشار 2001