Assessment of the DIXTAL DX-2710 automated oscillometric device for blood pressure measurement with the validation protocols of the British Hypertension Society (BHS) and the Association for the Advancement of Medical Instrumentation (AAMI).
نویسندگان
چکیده
OBJECTIVE To assess the Dixtal DX2710 automated oscillometric device used for blood pressure measurement according to the protocols of the BHS and the AAMI. METHODS Three blood pressure measurements were taken in 94 patients (53 females 15 to 80 years). The measurements were taken randomly by 2 observers trained to measure blood pressure with a mercury column device connected with an automated device. The device was classified according to the protocols of the BHS and AAMI. RESULT The mean of blood pressure levels obtained by the observers was 148+/-38/93+/-25 mmHg and that obtained with the device was 148+/-37/89+/-26 mmHg. Considering the differences between the measurements obtained by the observer and those obtained with the automated device according to the criteria of the BHS, the following classification was adopted: "A" for systolic pressure (69% of the differences < 5; 90% < 10; and 97% < 15 mmHg); and "B" for diastolic pressure (63% of the differences < 5; 83% < 10; and 93% < 15 mmHg). The mean and standard deviation of the differences were 0+/-6.27 mmHg for systolic pressure and 3.82+/-6.21 mmHg for diastolic pressure. CONCLUSION The Dixtal DX2710 device was approved according to the international recommendations.
منابع مشابه
A comparison of the British Hypertension Society and Association for the Advancement of Medical Instrumentation protocols for validating blood pressure measuring devices: can the two be reconciled?
BACKGROUND Experience with the original protocols of the Association for the Advancement of Medical Instrumentation (AAMI) and the British Hypertension Society (BHS) for validating blood pressure has provided valuable insight into the methodological problems associated with device validation and has influenced both the BHS and the AAMI in revising their protocols. OBJECTIVES To review the rev...
متن کاملQuality Control and Validation of Oscillometric Blood Pressure Measurements Taken During an Epidemiological Investigation
This study aims to validate blood pressure (BP) values measured by an oscillometric BP monitor and seek possible calibration methods if discrepancies exist. Noninvasive BP measurement outcomes were determined using an oscillometric BP monitor (Omron HBP-1300) versus a mercury sphygmomanometer (standard device). Two percent of subjects enrolled in an epidemiological investigation were systematic...
متن کاملProposals for simplifying the validation protocols of the British Hypertension Society and the Association for the Advancement of Medical Instrumentation.
Correspondence to Professor Eoin O’Brien, The Blood Pressure Unit, Beaumont Hospital, Dublin 9, Ireland. Tel : +353 1 809 21qO; fax: i-353 1 809 3005; e-mail: [email protected] Introduction Experience with the protocols of the British Hypertension Society (BHS) and the Association for the Advancement of Medical Instrumentation (AAMI) has provided valuable insight into the methodological problems a...
متن کاملThe 'International Protocol': more insight or more arithmetic?
It was with interest that we read the recent article concerning the ‘International Protocol’Fa newly proposed protocol for the validation of automated blood pressure measuring devices [1]. The objective of this new protocol was to simplify the most widely used protocols currently available, namely the British Hypertension Society (BHS) protocol 1993 and the protocol of the Association for the A...
متن کاملLimitations of current validation protocols for home blood pressure monitors for individual patients.
BACKGROUND Automatic blood pressure monitoring conducted at home is increasingly used in the diagnosis and management of hypertension. We assessed the adequacy of existing British Hypertension Society (BHS) and Association for the Advancement of Medical Instrumentation (AAMI) validation standards for automatic blood pressure monitoring devices. SUBJECT AND METHODS A theoretical study and an e...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
- Arquivos brasileiros de cardiologia
دوره 79 6 شماره
صفحات -
تاریخ انتشار 2002