Wear measurement of retrieved polyethylene ABG 1 cups by universal-type measuring microscope and X-ray methods.

نویسندگان

  • Jiri Gallo
  • Vitezslav Havranek
  • Ivana Cechova
  • Jana Zapletalova
چکیده

BACKGROUND Polyethylene wear is considered a most important part of periprosthetic osteolysis development. Thus, its measurement is central to contemporary orthopaedics. AIMS The aim of this paper was to compare the accuracy of three radiographic techniques for wear measurement. Secondly, the influence of the abduction angle of the cup on measurement accuracy was investigated. METHODS Wear was measured manually in 80 patients by a single observer according to the Livermore, Charnley, and Dorr description. A multi-component statistical analysis was used to test the hypothesis that the Livermore technique was superior. In vitro data obtained from a Universal-type measuring microscope served as a gold standard. RESULTS In vitro measurements showed an average linear wear of 0.363 mm per year (0.000-0.939, SD 0.241) with a corresponding volumetric wear rate of 161 mm3 per year (0-467, SD 118.2). The Livermore technique showed the least deviation from the optical reference standard and a superior position from the viewpoint of error analysis but the correlation coefficient was slightly less (r = 0.761) than for the Dorr and Charnley techniques (r = 0.795 and r = 0.778, respectively). In addition, the mean error of the Dorr method differed significantly from zero (p = 0.036). Overall, the Livermore technique was the most accurate method for polyethylene wear measurement regardless of the abduction angle of the cup. CONCLUSIONS The Livermore technique performed manually was more accurate than the Charnley and Dorr methods. Nevertheless, we consider the Dorr technique an adequate tool for day-to-day wear measurements, mainly due to its simplicity.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Influence of demographic, surgical and implant variables on wear rate and osteolysis in ABG I hip arthroplasty.

Periprosthetic osteolysis is associated with accelerated wear rates. The goal of this study was to investigate the influence of demographic and technical variables on wear rates and size of osteolytic lesions. Eighty retrieved ABG I prostheses were analyzed according to prospectively established criteria. There were 22 men and 58 women with an average age of 52 years (34-65) at the time of revi...

متن کامل

Validation of a new 2-D technique for radiographic wear measurement of cemented, highly cross-linked polyethylene acetabular cups

A new 2-D radiographic wear measurement system has been developed which enables the low wear of highly cross-linked polyethylene acetabular cups to be accurately and precisely measured from standard, pelvis radiographs. The software was validated using radiographic images of a measurement jig which could vary the cup orientation and simulate the effect of pelvic tilt/rotation. Wear was simulate...

متن کامل

Validation of a 3D CT method for measurement of linear wear of acetabular cups

BACKGROUND We evaluated the accuracy and repeatability of a 3D method for polyethylene acetabular cup wear measurements using computed tomography (CT). We propose that the method be used for clinical in vivo assessment of wear in acetabular cups. MATERIAL AND METHODS Ultra-high molecular weight polyethylene cups with a titanium mesh molded on the outside were subjected to wear using a hip sim...

متن کامل

In vivo and ex vivo measurement of polyethylene wear in total hip arthroplasty

BACKGROUND Determination of the amount of wear in a polyethylene liner following total hip arthroplasty (THA) is important for both the clinical care of individual patients and the development of new types of liners. PATIENTS AND METHODS We measured in vivo wear of the polyethylene liner using computed tomography (CT) (obtained in the course of regular clinical care) and compared it to coordi...

متن کامل

Quantification of Wear and Deformation in Different Configurations of Polyethylene Acetabular Cups Using Micro X-ray Computed Tomography

Wear is currently quantified as mass loss of the bearing materials measured using gravimetric methods. However, this method does not provide other information, such as volumetric loss or surface deviation. In this work, we validated a technique to quantify polyethylene wear in three different batches of ultrahigh-molecular-polyethylene acetabular cups used for hip implants using nondestructive ...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:
  • Biomedical papers of the Medical Faculty of the University Palacky, Olomouc, Czechoslovakia

دوره 150 2  شماره 

صفحات  -

تاریخ انتشار 2006