Sensitive determination of impurities in achiral pharmaceuticals by supercritical fluid chromatography using the Agilent 1260 Infinity Analytical SFC System
نویسندگان
چکیده
Supercritical Fluid Chromatography is an alternative and complementary method to HPLC for the determination of impurities in pharmaceutical products. This study evaluated the performance of the Agilent 1260 Infinity Analytical SFC system for the analysis of thiourea in two model compounds. The study included both isocratic and modifier gradient modes. The resulting data showed excellent linearity and repeatability. Retention time repeatability was better than 0.1%. Limits of detection for the impurity determination were below 0.01% w/w relative to the main compound, and the RSDs of peak area were below 1% at the 0.5% concentration level and below 5% at the 0.05% concentration level in all cases. These values are well within acceptable method validation ranges. Authors Melissa N. Dunkle, Alberto dos Santos Pereira, Frank David, Pat Sandra Research Institute for Chromatography Kennedypark 26, 8500 Kortrijk Belgium Application Note
منابع مشابه
Determination of Enantiomeric Excess of Metolachlor from Chiral Synthesis using the Agilent 1260 Infinity Analytical SFC System
This Application Note describes the development of a method for the separation of four stereoisomers, which were derived from a compound with two steric centers. The method was developed using an Agilent 1260 Infinity Analytical Supercritical Fluid Chromatography (SFC) System with Agilent ChemStation Method Scouting Wizard software. The developed method was used to compare the ratio of stereois...
متن کاملDetermination of Enantiomeric Excess of Metolachlor from Chiral Synthesis using the Agilent 1260 Infi nity Analytical SFC System
This Application Note describes the development of a method for the separation of four stereoisomers, which were derived from a compound with two steric centers. The method was developed using an Agilent 1260 Infi nity Analytical Supercritical Fluid Chromatography (SFC) solution with Agilent ChemStation Method Scouting Wizard software. The developed method was used to compare the ratio of stere...
متن کاملFeed Speed and Overfeed Volume New Parameters for Injection in Supercritical Fluid Chromatography
This Technical Overview demonstrates and discusses the influence of the injection parameters Feed Speed and Overfeed Volume on chromatographic separation. These parameters were introduced with the Agilent 1260 Infinity II Supercritical Fluid Chromatography (SFC) Multisampler. The influence on the isocratic separation of closely eluting compounds, shown for different injection volumes commonly u...
متن کاملSuperior sensitivity by seamlessly interfacing to the Agilent 6100
The applicability of Supercritical Fluid Chromatography (SFC) can be substantially enhanced by coupling SFC to Mass Spectrometric detection (MSD). In this Technical Overview, two instrument configurations are described that can be used to connect the Agilent 1260 Infinity Analytical SFC to a mass spectrometer either using an atmospheric pressure chemical ionization (APCI) or an electrospray (ES...
متن کاملAgilent Application Solution Transfer of a USP method for prednisolone from normal phase HPLC to SFC using the Agilent 1260 Infinity Hybrid SFC/UHPLC System
Normal phase liquid chromatography (LC) methods often have long run times and involve environmentally toxic/costly solvents. Supercritical fluid chromatography (SFC) methods on the other hand are faster, inexpensive, and eco-friendly. SFC involves the use of low viscosity supercritical carbon dioxide that can be operated at flow rates up to 3x higher than LC without losing separation efficiency...