Fatigue fracture of the femoral component in a mobile bearing knee prosthesis.

نویسنده

  • Roger Lemaire
چکیده

Stress fracture of the femoral component is a rare complication following total knee arthroplasty. We report one such complication which occurred 6.5 years after primary implantation of an uncemented meniscal Low Contact Stress (LCS) prosthesis. The fracture affected the medial condyle of the femoral component, and a large osteolytic lesion was present in the underlying femoral condyle. The patient underwent revision arthroplasty with a cemented constrained condylar knee, which is still in place with a satisfactory result 12 years later. This complication appears to be very rare, although a number of cases probably go unreported. From data available in literature, it appears that stress fractures of the femoral component have predominantly affected the medial condyle, following uncemented implantation of fixed-bearing as well as mobile-bearing knees. Different mechanisms may be involved: specific design features of the implant, failure of bone ingrowth in uncemented components, and osteolysis from particulate wear debris.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Polyethylene Bearing Stress Fractures after Mobile-bearing Cruciate-retaining Total Knee Arthroplasty: Case Reports

Some reported bearing fractures are due to entrapment of the bearing between the femoral component and posterior edge of the tibial base plate in unstable meniscal –bearing design TKA. However, so far no cases of bearing fracture has been reported for the rotating platform or anteroposterior gliding designs of mobile bearing TKA. The authors describe two cases of polyethylene stress fracture af...

متن کامل

Nontraumatic Fracture of the Femoral Condylar Prosthesis in a Total Knee Arthroplasty Leading to Mechanical Failure

This paper reports a case of fatigue fracture of the femoral component in a cruciate-retaining cemented total knee arthroplasty (TKA). A 64-year-old man had undergone a primary TKA for osteoarthritis 10 years previously at another institution using the PFC-Sigma prosthesis. The patient recovered fully and was back to his regular activities. He presented with a history of sudden onset pain and l...

متن کامل

Stress Fracture of the Lateral Femoral Condyle afterTotal Knee Arthroplasty

Background: This is to assess one of the rare complications after total knee replacement and to assess risk factors of failure.Methods: 11 patients with varus knee and an average age of 67 years underwent TKA between 2005 and 2013. All patients returned with a sudden sharp knee pain, disability to walk and significant decrease in ROM about 4 to 8 weeks after surgery. Radiographic examination re...

متن کامل

Fracture of mobile unicompartmental knee bearings: a parametric finite element study.

Cases of fractured mobile unicompartmental knee bearings have recently been reported. The purpose of this study was to understand the mechanics behind these fractures and to examine the influence of different design modifications. A parametric finite element model was used to examine the influence of different geometrical factors on the stresses within the bearing. Crack initiation occurred cli...

متن کامل

Nontraumatic tibial polyethylene insert cone fracture in mobile-bearing posterior-stabilized total knee arthroplasty

A 72-year-old male patient underwent mobile-bearing posterior-stabilized total knee arthroplasty for osteoarthritis. He experienced a nontraumatic polyethylene tibial insert cone fracture 27 months after surgery. Scanning electron microscopy of the fracture surface of the tibial insert cone suggested progress of ductile breaking from the posterior toward the anterior of the cone due to repeated...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:
  • Acta orthopaedica Belgica

دوره 76 2  شماره 

صفحات  -

تاریخ انتشار 2010