Low contact force and force–time integral predict early recovery and dormant conduction revealed by adenosine after pulmonary vein isolation
نویسندگان
چکیده
AIM After pulmonary vein isolation (PVI), dormant conduction (DC) is present in at least one vein in a substantial number of patients. The present study seeks to determine whether there is a relationship between poor contact forces (CF) and the presence of DC after PVI. METHODS AND RESULTS This prospective, operator-blinded, non-randomized dual-centre trial enrolled 34 consecutive patients with paroxysmal atrial fibrillation who were candidates for PVI. Radiofrequency (RF) energy was delivered by using an irrigated-tip force-sensing ablation catheter (Tacticath, St Jude Medical) at pre-defined target power. The operators were blinded to the CF data at all times. A total of 1476 RF applications were delivered in 743 pre-defined PV segments. For each application, the precise location of the catheter was registered and the following data were extracted from the Tacisys unit: application duration, minimum contact force, maximum contact force, average contact force (CF), and force-time integral (FTI). Sixty minutes after PVI, spontaneous early recovery (ER) of the left atrium (LA) to PV conduction was evaluated. In the absence of ER, the presence of a DC was evaluated by using intravenous adenosine (ATP). In the 34 patients recruited (23 males; mean age: 62 ± 9 years), all PVs were successfully isolated. At the end of the 60 min waiting period, 22 patients demonstrated at least one spontaneous ER or DC under ATP. The mean CF and FTI per PV segment differed significantly among the different veins but the sites of ER and DC were evenly distributed. However, both the minimum, the first and the mean CF and FTI per PV segment were significantly lower in the PV segments presenting either ER or DC as compared with those without ER or DC (mean CF: 4.9 ± 4.8 vs. 12.2 ± 1.65 g and mean FTI: 297 ± 291 vs. 860 ± 81 g s, P < 0.001 for both). Using multivariate analysis, both the mean CF and the FTI per lesion remained significantly associated with the risk of ER or DC. Moreover, a CF < 5 g per PV segment predicted ER+ and DC+ with a sensitivity of 71% and specificity of 82%. In contrast, ER and DC were very unlikely if RF application was performed with a mean CF > 10 g (negative predictive value: 98.7%). CONCLUSION Both a low CF and a low FTI are associated with the ER of the PVI and DC after PVI.
منابع مشابه
Optimal Force–Time Integral for Pulmonary Vein Isolation According to Anatomical Wall Thickness Under the Ablation Line
BACKGROUND Low contact force and force-time integral (FTI) during catheter ablation are associated with ineffective lesion formation, whereas excessively high contact force and FTI may increase the risk of complications. We sought to evaluate the optimal FTI for pulmonary vein (PV) isolation based on atrial wall thickness under the ablation line. METHODS AND RESULTS Contact force parameters a...
متن کاملAdenosine following pulmonary vein isolation to target dormant conduction elimination (ADVICE): methods and rationale.
BACKGROUND Pulmonary vein (PV) isolation (PVI) has emerged as an effective therapy for paroxysmal atrial fibrillation (AF). However, AF recurs in up to 50% of patients, generally because of recovery of PV conduction. Adenosine given during the initial procedure may reveal dormant PV conduction, thereby identifying the need for additional ablation, leading to improved outcomes. The Adenosine Fol...
متن کاملClinical implication of adenosine test at repeat atrial fibrillation ablation procedure: the importance of detecting dormant thoracic vein conduction.
BACKGROUND Pulmonary vein reconnection after electrical isolation is commonly observed in the context of atrial fibrillation ablation and is associated with recurrent atrial tachyarrhythmias. Adenosine test was been performed to identify acute dormant conduction immediately after pulmonary vein isolation at index procedure. However, the utility of adenosine test at repeat procedure has not been...
متن کاملThe Role of Adenosine in Pulmonary Vein Isolation: A Critical Review
The cornerstone of atrial fibrillation (AF) ablation is pulmonary vein isolation (PVI), which can be achieved in more than 95% of patients at the end of the procedure. However, AF recurrence rates remain high and are related to recovery of PV conduction. Adenosine testing is used to unmask dormant pulmonary vein conduction (DC). The aim of this study is to review the available literature addres...
متن کاملContact Force in Atrial Fibrillation: Role of Atrial Rhythm and Ventricular Contractions: Co-Force Atrial Fibrillation Study.
BACKGROUND In an experimental model, variable and intermittent contact force (CF) resulted in a significant decrease in lesion volume. In humans, variability of CF during pulmonary vein isolation has not been characterized. METHODS AND RESULTS In 20 consecutive patients undergoing CF-guided circumferential pulmonary vein isolation, 914 radiofrequency applications (530 in sinus rhythm and 384 ...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
دوره 17 شماره
صفحات -