Comparison of Thyroglobulin Measurements Using Three Different Immunoassay Kits: A BRAMHS Tg-Plus RIA Kit, a BRAMHS hTg Sensitive Kryptor Kit, and a Beckman Coulter ACCESS Immunoassay Kit

نویسندگان

  • Mijin Kim
  • Min Ji Jeon
  • Won Gu Kim
  • Jong Jin Lee
  • Jin Sook Ryu
  • Eun Jung Cho
  • Dae Hyun Ko
  • Woochang Lee
  • Sail Chun
  • Won Ki Min
  • Tae Yong Kim
  • Young Kee Shong
  • Won Bae Kim
چکیده

BACKGROUND Second-generation thyroglobulin immunometric assays (Tg-IMAs) have been developed with improved sensitivity. Our aim was to compare the diagnostic value of Tg-IMA measurements using a Kryptor (BRAHMS AG) kit (Tg-K) and an ACCESS (Beckman Coulter) kit (Tg-A) with that of the first-generation Tg measurement using a Tg-plus (BRAHMS AG) kit (Tg+). METHODS We enrolled 82 differentiated thyroid cancer patients who underwent total thyroidectomy with radioactive iodine remnant ablation and who underwent diagnostic whole body scan using recombinant human thyroid stimulating hormone (rhTSH). The Tg+, Tg-K, and Tg-A were measured before rhTSH administration during levothyroxine treatment (suppressed Tg) from the same sample. Serum Tg+ was measured after rhTSH stimulation (stimulated Tg). RESULTS Suppressed Tg+ was more significantly correlated with suppressed Tg-K (R²=0.919, P<0.001) than with suppressed Tg-A (R²=0.536, P<0.001). The optimal cut-off values of suppressed Tg+, Tg-K, and Tg-A for predicting stimulated Tg+ of 1 ng/mL were 0.3, 0.2, and 0.2 ng/mL, respectively. The sensitivity, specificity, and accuracy of suppressed Tg+ were 67%, 100%, and 90%, respectively; those of suppressed Tg-K were 83%, 90%, and 88%; those of suppressed Tg-A were 96%, 82%, and 87%, respectively. The positive predictive and negative predictive values of Tg+ were 100% and 87%, respectively; those of Tg-K were 79% and 92%; and those of Tg-A were 73% and 98%. CONCLUSION We could not clearly demonstrate which kit had better diagnostic performance after comparison of first-generation Tg measurements with Tg-IMA measurements. Also, there were kit-to-kit variations between Tg-IMA kits. Suppressed Tg measured by Tg-IMA was insufficient to completely substitute for a stimulated Tg measurement.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Intercomparison of seven radioimmunoassay kits and a fluorescence polarization immunoassay kit for digoxin.

Samples from 33 patients being treated with digoxin and three concentrations of control material were analyzed for this drug by use of seven radioimmunoassay (RIA) kits and an automated fluorescence polarization immunoassay (FPIA) system. CVs ranged from 2.4 to 6.4% for the FPIA system and from 4.6 to 7.4% for two RIA methods. In the analysis of between-method variability of RIA kits, CVs range...

متن کامل

Circulating soluble apo 2L, soluble OX-2: a possible screening biomarker for stage-4 ovarian, endometrial CA

Methods Ascites and sera samples from patients without known malignancy at the admission were collected. There were 14 stage-4 breast cancer (BC), 17 stage-4 ovarian cancer (OC) and 19 stage-4 endometrial cancer (EC) diagnosed later on. Control groups consisted of benign peritoneal fluids (n=53) and sera samples (n=25) from healthy subjects. Concentrations of sCD200 in the serum samples were qu...

متن کامل

Decreasing Quality of the New Generations of Anti-Müllerian Hormone Assays

Anti-Müllerian hormone (AMH) measurements are widely used to optimize the stimulation protocols. First generation AMH kits correlated well with ovarian reserve and response to stimulation. In the present study we aimed to asses if the new generation kits share the same accurate correlations. Retrospective data were collected from 8323 blood samples. For comparison we used Immunotech I generatio...

متن کامل

Limitations of direct estradiol and testosterone immunoassay kits.

Estradiol (E2) and testosterone (T) are biologically active hormones that serve as important diagnostic markers in serum of premenopausal and postmenopausal women and in men. These hormones are measured frequently by immunoassay in clinical laboratories and the test results are used in the diagnosis and treatment of patients. For measuring the hormones by immunoassay, most laboratories utilize ...

متن کامل

Positive interference in Radioimmunoassay determination of serum total T3

Introduction: Determination of thyrotropin, total and free thyroxine and triiodothyronine are widely used as diagnostic methods for thyroid function evaluation. There have been numerous reports of interferences in thyroid hormone immunoassays. Herein, a prominent positive interference is reported for a radioimmunoassay kit of total triiodothyronine. Materials and Methods: In this study, a total...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره 31  شماره 

صفحات  -

تاریخ انتشار 2016