Large discrepancies between planned and actually delivered dose in IMRT of head and neck cancer. A case report.
نویسندگان
چکیده
The case is reported of a patient with locally recurrent carcinoma of the tongue treated with intensity-modulated radiotherapy (IMRT) (simultaneous integrated boost) plus concurrent chemotherapy, who during the third week of radiotherapy developed grade 3 mucositis. Treatment was interrupted for 10 days until significant resolution of the symptoms. At the time of treatment resumption the patient showed 8% weight loss, and in vivo portal dose verification revealed large discrepancies between the computed and measured doses. A new CT scan showed marked tumor shrinkage and modifications to the critical structures. The comparison between the original plan and the hybrid IMRT showed a minimal dose increase in the new target volumes and a marked dose increase in the organs at risk. This case confirms the need for a robust quality assurance program when using IMRT, the feasibility and efficacy of in vivo dosimetry to detect significant discrepancies between planned and delivered dose, and the need to combine IMRT with 4-dimensional radiotherapy, at least for head and neck cancer.
منابع مشابه
Dose Distribution Analysis of Rapid Arc and Intensity Modulated Radiotherapy Plan in Head and Neck Cancer
Introduction: Globally, intensity-modulated radiation therapy (IMRT) is considered as highly precise and accurate method of radiotherapy planning. This technique amplifies spatial dose distribution conformity by modulating the intensity of radiation beams in each sub-volume. Additionally, it can reduces the dose to surrounding critical organs and deliver the planned dose to targets with the nom...
متن کاملAssessment of a 2D EPID-based Dosimetry Algorithm for Pre-treatment and In-vivo Midplane Dose Verification
Introduction: The use of electronic portal imaging devices (EPIDs) is a method for the dosimetric verification of radiotherapy plans both pretreatment and in-vivo. The aim of this study was to test a 2D EPID-based dosimetry algorithm for dose verification of some plans inside a homogenous and anthropomorphic phantom and in-vivo, as well. Materials and Methods: </strong...
متن کاملComprehensive end-to-end test for intensity-modulated radiation therapy for nasopharyngeal carcinoma using an anthropomorphic phantom and EBT3 film
Background: In head and neck radiotherapy, immobilization devices can affect dose delivery. In this study, a comprehensive end-to-end test was developed to evaluate the accuracy of radiotherapy treatment. Materials and Methods: An Alderson Radiation Therapy (ART) anthropomorphic phantom with EBT3 film was used to mimic the actual patient treatment process. Ten patients treated for nasopharyngea...
متن کاملInvestigation of VMAT Algorithms and Dosimetry.
Purpose: Planning and dosimetry of different VMAT algorithms (SmartArc, Ergo++, Autobeam) is compared with IMRT for Head and Neck Cancer patients. Modelling was performed to rule out the causes of discrepancies between planned and delivered dose. Methods: Five HNC patients previously treated with IMRT were re-planned with SmartArc (SA), Ergo++ and Autobeam. Plans were compared with each other a...
متن کاملLhermitte's Sign Developing after IMRT for Head and Neck Cancer
Background. Lhermitte's sign (LS) is a benign form of myelopathy with neck flexion producing an unpleasant electric-shock sensation radiating down the extremities. Although rare, it can occur after head and neck radiotherapy. Results. We report a case of Lhermitte's developing after curative intensity-modulated radiotherapy (IMRT) for a patient with locoregionally advanced oropharyngeal cancer....
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
- Tumori
دوره 93 3 شماره
صفحات -
تاریخ انتشار 2007