Debonding of the acetabular porous coating in hip resurfacing arthroplasty. A case report.
نویسندگان
چکیده
This report presents the case of a patient in whom hip resurfacing arthroplasty failed because of loosening of the acetabular shell secondary to debonding of the surface coating. During normal function, there will inevitably be cyclic shear stress between the coating and the surrounding bone. This might result in fatigue failure of the bond. In spite of increasing numbers of porous coated acetabular shells being implanted, little is known about the fatigue performance of the interface between coating and implant.
منابع مشابه
Hip Arthroplasty and its Revision in a Child: Case Report and Literature Review
Juvenile idiopathic arthritis is the leading cause of hip replacement in young children. However, arthroplasty in this population is challenging with several concerns about quality of the growing bone, young age for revision surgery, and difficulties in potential several revisions. In this study we introduce a case of a 12-year old who is one of the youngest patients to undergo revision hip art...
متن کاملEarly Clinical Outcomes with a 3-D Porous Titanium Acetabular Cup
Cementless acetabular components are associated with a significant incidence of polyethylene wear and secondary osteolysis. 3-dimentional (3-D) porous coating and enhanced shell to a liner fixation are expected to reduce the polyethylene wear and to increase the longevity of an acetabular cup. The authors report the early clinical outcome with a cementless acetabular cup, Regenerex Ringloc+ (Zi...
متن کاملA randomised study comparing resection of acetabular bone at resurfacing and total hip replacement.
We have undertaken a prospective, randomised study to compare conservation of acetabular bone after total hip replacement and resurfacing arthroplasty of the hip. We randomly assigned 210 hips to one of the two treatment groups. Uncemented, press-fit acetabular components were used for both. No significant difference was found in the mean diameter of acetabular implant inserted in the groups (5...
متن کاملPoor outcome of revised resurfacing hip arthroplasty
BACKGROUND AND PURPOSE Recent years have seen a rapid increase in the use of resurfacing hip arthroplasty despite the lack of literature on the long-term outcome. In particular, there is little evidence regarding the outcome of revisions of primary resurfacing. The purpose of this analysis was to examine the survivorship of primary resurfacing hip arthroplasties that have been revised. PATIEN...
متن کاملComparison of acetabular reamings during hip resurfacing versus uncemented total hip arthroplasty.
PURPOSE To compare the quantity of bone removed from the acetabulum during resurfacing hip arthroplasty versus uncemented total hip arthroplasty (THA). METHODS 62 consecutive patients with osteoarthritis of the hip were prospectively studied. 24 men and 7 women aged 40 to 86 (mean, 59) years underwent Birmingham hip resurfacing. 13 men and 18 women aged 34 to 88 (mean, 61) years underwent unc...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
- Hip international : the journal of clinical and experimental research on hip pathology and therapy
دوره 22 2 شماره
صفحات -
تاریخ انتشار 2011