Verification of junction dose between VMAT arcs of total body irradiation using a Sun Nuclear ArcCHECK phantom

نویسندگان

  • Colm T. Morrison
  • Kirsty L. Symons
  • Simon J. Woodings
  • Michael J. House
چکیده

A volumetric modulated arc therapy (VMAT) approach to total body irradiation (TBI) has recently been introduced at our institution. The planning target volume (PTV) is divided into separate sub-volumes, each being treated with 2 arcs with their own isocentre. Pre-treatment quality assurance of beams is performed on a Sun Nuclear ArcCHECK diode array. Measurement of junction regions between VMAT arcs with separate isocentres has previously been performed with point dose ionization chamber measurements, or with films. Translations of the ArcCHECK with respect to a known distance between the adjacent isocentres of two arcs, which are repeated with the ArcCHECK in an inverted position, allows the recording of a junction dose map. A 3%/3 mm global gamma analysis (10% threshold) pass rate for arc junctions were comparable to their component arcs. Dose maps of junction regions between adjacent arcs with different isocentres can be readily measured on a Sun Nuclear ArcCHECK diode array.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Volumetric modulated arc therapy for total body irradiation: A feasibility study using Pinnacle3 treatment planning system and Elekta Agility™ linac

A study was undertaken to explore the use of volumetric modulated arc therapy (VMAT) for total body irradiation (TBI). Five patient plans were created in Pinnacle3 using nine 6 MV photon dynamic arcs. A dose of 12 Gy in six fractions was prescribed. The planning target volume (PTV) was split into four subsections for the head, chest, abdomen, and pelvis. The head and chest beams were optimized ...

متن کامل

Commissioning and comprehensive evaluation of the ArcCHECK cylindrical diode array for VMAT pretreatment delivery QA

We present commissioning and comprehensive evaluation for ArcCHECK as a QA equipment for volumetric-modulated arc therapy (VMAT), using the 6 MV photon beam with and without the flattening filter, and the SNC patient software (version 6.2). In addition to commissioning involving absolute dose calibration, array calibration, and PMMA density verification, ArcCHECK was evaluated for its response ...

متن کامل

SU-F-T-311: Comparison of Measurements From Two Diode Array QA Devices to Deep Point Dose Measurements with Two Treatment Planning Model Settings for Brain VMAT SRT Patient-Specific QA.

PURPOSE To investigate the sensitivity of traditional gamma-index-based analysis using diode arrays to detect significant dosimetric errors in patientspecific measurements for VMAT brain SRT plans. METHODS Quality assurance measurements were performed using an ion chamber and two different diode array devices for four brain SRT treatment plans. All cases were planned with Eclipse AAA RapidArc...

متن کامل

Patient-specific quality assurance of RapidArc treatments: Portal prediction dosimetry compared with phantom studies

Purpose: To validate a locally fabricated phantom of Imatrixx-2D Array by comparing its results with ArcCheck phantom and comparing portal dosimetry measurements with the two phantom studies. Materials and Methods: Electronic Portal Imaging Devices and Epiqa software were used for portal dosimetry. An Imatrixx-2D array with a locally fabricated phantom and ArcCheck cylindrical phantom were used...

متن کامل

Purpose: to validate the clinical effectiveness of Control Point Analysis (CPA) software and establish a comprehensive quality assurance (QA) program for Volumetric Modulated Arc Therapy (VMAT) treatment. Methods: The comparison and gamma analysis of ArcCHECKTM measurement and treatment planning

ID: 19178 Title: Establishing Comprehensive Quality Assurance Program for Volumetric Modulated Arc Therapy -Clinical Study Using Control Point Analysis Software Purpose: to validate the clinical effectiveness of Control Point Analysis (CPA) software and establish a comprehensive quality assurance (QA) program for Volumetric Modulated Arc Therapy (VMAT) treatment. Methods: The comparison and gam...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره 18  شماره 

صفحات  -

تاریخ انتشار 2017