Prevalence of Calibration Errors in Goldmann Applanation Tonometers
نویسندگان
چکیده
PURPOSE To determine the prevalence of calibration errors in Goldmann applanation tonometers at Farabi Eye Hospital. METHODS This cross-sectional study was performed on all tonometers in use at Farabi Eye Hospital. All Haag-Streit Goldmann applanation tonometers were checked according to the manufacturer's method by two independent observers and by a third observer in case of mismatched results. Calibration errors were classified into 6 categories of ±0.5, ±1, ±1.5, ±2, ±2.5 and more than ±2.5 mmHg. RESULTS Overall, 43 Goldmann tonometers were evaluated. There were 3 (7%), 10 (24.3%), 16 (38.3%), 24 (56.9%), 31 (72.1%) and 12 (27.9%) tonometers within calibration errors of ±0.5, ±1, ±1.5, ±2, ±2.5 and more than ±2.5 mmHg respectively. CONCLUSION Goldmann tonometers were not within the manufacturer's recommended range (±0.5 mmHg) in 93%, and not within the acceptable range of ±2.5 mmHg in 28% of checked devices. Further study is needed to demonstrate the correlation between calibration errors and clinical errors.
منابع مشابه
Have You Tested Your Goldmann Tonometer Today?
Intraocular pressure (IOP) is the only known modifiable risk factor for glaucoma. Besides being a cornerstone in glaucoma management, accurate measurement of IOP forms an important component of routine ophthalmic evaluation. Despite the innovation of various newer tonometers, Goldmann applanation tonometer is still considered the Gold standard for IOP measurement. However, there are several rep...
متن کاملTonometer calibration in Brasília, Brazil.
PURPOSE To determine calibration errors of Goldmann applanation tonometers in ophthalmic clinics of Brasília, Brazil, and correlate the findings with variables related to tonometers model and utilization. METHODS Tonometers from ophthalmic clinics in Brasília, Brazil, were checked for calibration errors. A standard Goldmann applanation tonometer checking tool was used to asses the calibration...
متن کاملGoldmann applanation tonometry in the conscious rat.
PURPOSE To determine whether the Goldmann applanation tonometer can be modified to measure intraocular pressure (IOP) in the conscious rat. METHODS In anesthetized rats Goldmann tonometers were tested that had reduced biprism angles in the applanating tips and reduced weights in the tonometer body from those used in humans and species with similar size eyes. Tonometers with tips with biprism ...
متن کاملAuthor’s response to reviews Title: Comparison of Intraocular Pressure as Measured by Three Different Non-contact Tonometers and Goldmann Applanation Tonometer for Non-glaucomatous Subjects Authors:
Title: Comparison of Intraocular Pressure as Measured by Three Different Non-contact Tonometers and Goldmann Applanation Tonometer for Non-glaucomatous Subjects Authors: Seung Pil Bang ([email protected]) Chong Eun Lee ([email protected]) Yu Cheol Kim ([email protected]) Version: 3 Date: 27 Jul 2017 Author’s response to reviews: Revision Letter Manuscript Number BOPH-D-17-00287 “Comparison of Int...
متن کاملRe-evaluation of the Schiotz tonometer calibration.
The calibration scale for the Schi0tz tonometers toas tested by comparing Goldmann applanation and Schi0tz measurements on the eyes of 906 patients. The study differed from several previous studies in that both the applanation and Schi0tz measurements were made with the subjects lying on their backs, thus eliminating errors due to change in intraocular pressure as the subject changed from one p...