First-pass perfusion CMR with reduced dark-rim artifact and instantaneous image reconstruction using optimized cartesian sampling and apodization
نویسندگان
چکیده
Background Subendocardial dark-rim artifact (DRA) continues to be a major issue that limits the diagnostic performance of first pass perfusion (FPP) CMR. Non-Cartesian approaches such as radial or spiral acquisition have been proposed to minimize DRAs. Among these approaches, those that can operate without the need for breath-holding typically require a time-consuming offline image reconstruction procedure, which limits their clinical accessibility. We propose a free-breathing DRA-reduced FPP scheme with instant image reconstruction on the scanner.
منابع مشابه
Towards elimination of the dark-rim artifact in first-pass myocardial perfusion MRI: removing Gibbs ringing effects using optimized radial imaging.
PURPOSE Subendocardial dark-rim artifacts (DRAs) remain a major concern in first-pass perfusion (FPP) myocardial MRI and may lower the diagnostic accuracy for detection of ischemia. A major source of DRAs is the "Gibbs ringing" effect. We propose an optimized radial acquisition strategy aimed at eliminating ringing-induced DRAs in FPP. THEORY AND METHODS By studying the underlying point sprea...
متن کاملSimulation of Motion-Induced Dark-Rim Artifacts for Cartesian and Spiral Pulse Sequences
Introduction: First-pass MR perfusion images may be corrupted by dark-rim artifacts, which are at least in part a consequence of myocardial motion during data acquisition [1]. We have previously shown that the non-linear time-varying displacements of the myocardium resulting from cardiac contraction result in dark-rim artifacts that have a different appearance in images obtained at different ph...
متن کاملEliminating dark-rim artifacts in first-pass myocardial perfusion imaging
Background We demonstrate that projection imaging significantly reduces the prevalence and spatial extent of subendocardial dark-rim artifacts (DRAs) in first-pass perfusion (FPP) myocardial MR, compared to conventional Cartesian techniques. A major cause of DRAs, which remain a major concern in FPP imaging, is known to be the socalled Gibbs ringing (truncation) phenomenon [1-3]. Radial k-space...
متن کاملEffect of Improving Spatial or Temporal Resolution on Image Quality and Quantitative Perfusion Assessment with k-t SENSE Acceleration in First-Pass CMR Myocardial Perfusion Imaging
k-t Sensitivity-encoded (k-t SENSE) acceleration has been used to improve spatial resolution, temporal resolution, and slice coverage in first-pass cardiac magnetic resonance myocardial perfusion imaging. This study compares the effect of investing the speed-up afforded by k-t SENSE acceleration in spatial or temporal resolution. Ten healthy volunteers underwent adenosine stress myocardial perf...
متن کاملAdenosine stress CMR with spiral pulse sequences accurately detect CAD
Background Adenosine stress perfusion imaging with CMR has been limited by motion-induced dark-rim artifacts, which may be mistaken for true perfusion abnormalities. We have previously demonstrated that spiral pulse sequences can produce high quality first-pass perfusion images.[1] We have further improved this spiral technique using high-resolution variable density spiral trajectories and a no...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
دوره 17 شماره
صفحات -
تاریخ انتشار 2015