Portal dosimetry for pretreatment verification of IMRT plan: a comparison with 2D ion chamber array

نویسندگان

  • Dayananda Shamurailatpam Sharma
  • Vaibav Mhatre
  • Malhotra Heigrujam
  • Kaustav Talapatra
  • Suman Mallik
چکیده

Portal dosimetry (PD) was performed for 181 fields from 14 IMRT plans of various clinical sites at gantry zero and source-to-detector distance (SDD) of 100 cm. PD was realized using aSi1000 electronic portal imaging device (EPID) and portal dose prediction (PDP) algorithm implemented in Eclipse treatment planning system (TPS). Agreement of PDP predicted and EPID measured photon fluence/dose distribution were evaluated using gamma (γ) index set at 3% at 3 mm distance to point agreement (DTA). Three gamma scaling parameters, maximum γ (γ(max)), average γ (γ(avg)) and percentage of points with γ ≤ 1 (γ% ≤ 1) were estimated for each field. An independent measurement was carried out using MatriXX 2D ion chamber array with detector plane at 100 cm and γ(max), γ(avg) and γ% ≤ 1 were estimated using OmniPro IMRT analyzing software. Effect of extended SDD and gantry rotation on portal dosimetry outcome was also investigated for another 45 IMRT fields. PDP predicted and EPID measured photon fluence agrees well with overall mean values of γ(max), γ(avg) and γ% ≤ 1 at 2.02, 0.24 and 99.43%, respectively. γ(max) value was lower in 15 MV compared to 6 MV IMRT plan. Independent verification using MatriXX showed comparable overall mean values of γ(avg) and γ% ≤ 1 at 0.25 and 99.80%. However, in all plans, MatriXX showed significantly lower γ(max) (p < 0.05) with an overall mean value of 1.35. In portal dosimetry, compared to gamma values at 100 cm SDD, γ(max), γ(avg) and γ% ≤ 1 values improve from a mean of 0.16, 0.03 and 0.26 at 110 cm SDD to 0.35, 0.05 and 0.29 at 140 cm SDD. PD outcome was independent of gantry rotation. In conclusion, both MatriXX 2D ion chamber array and portal dosimetry showed comparable results and can be use as an alternative to each other for relative photon fluence verification.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Evaluation of a fast method of EPID-based dosimetry for IMRT and Comparison with 3D EPID-based dosimetry system using conventional two- and three-dimensional detectors for VMAT

Introduction: Electronic portal imaging devices (EPIDs) could potentially be useful for intensity-modulated radiation therapy (IMRT) and VMAT QA. The data density, high resolution, large active area, and efficiency of the MV EPID make it an attractive option. However, EPIDs were designed as imaging devices, not dosimeters, and as a result they do not inherently measure dose in ...

متن کامل

Assessment of a 2D EPID-based Dosimetry Algorithm for Pre-treatment and In-vivo Midplane Dose Verification

Introduction: The use of electronic portal imaging devices (EPIDs) is a method for the dosimetric verification of radiotherapy plans both pretreatment and in-vivo. The aim of this study was to test a 2D EPID-based dosimetry algorithm for dose verification of some plans inside a homogenous and anthropomorphic phantom and in-vivo, as well. Materials and Methods: </strong...

متن کامل

Patient Specific IMRT Pre-Treatment Verification Using A Commercial Portal Dosimetry Software

Introduction: Electronic portal imaging devices (EPIDs) are extensively used for dosimetric information of pre-treatment field verification and in-vivo dosimetry for Intensity Modulated Radiotherapy (IMRT). In the present work, we have implemented the newly developed portal dosimetry software using independent dose prediction algorithm EPIDose and evaluated this new tool for the pretreatment IM...

متن کامل

Camera based EPID dosimetric verification of radiation treatments

Introduction: Tumor control and treatment quality depend on accuracy of delivered dose. Various methods have been studied for treatment dosimetry verification. Among these methods, the use of electronic portal imaging device (EPID) is highly considered. Many efforts are underway to improve the results of this context.   Materials and Methods: In this study, ...

متن کامل

Dosimetric study of 2D ion chamber array matrix for the modern radiotherapy treatment verification

Intensity-modulated radiotherapy treatment demands stringent quality assurance and accurate dose determination for delivery of highly conformal dose to the patients. Generally 3D dose distributions obtained from a treatment planning system have to be verified by dosimetric methods. Mainly, a comparison of two-dimensional calculated and measured data in several coplanar planes is performed. In p...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره 11  شماره 

صفحات  -

تاریخ انتشار 2010