Isoflurane cracks the polycarbonate connector of extra-corporeal circuit -A case report-

نویسندگان

  • Hyung Sun Lim
  • Sung Hun Cho
  • Deok Kyu Kim
  • Sang Kyi Lee
  • Seong Hoon Ko
چکیده

Cardiopulmonary bypass (CPB) is widely used for cardiac surgery by virtue of its proven safety over the course of its use during the past half century. Even though perfusion is safer, incidents still occur. During the repair of a ventricular-septal defect in an 11-month-old infant, we experienced a critical incident related to the potential hazardous effect of volatile anesthetics on the polycarbonate connector of extra-corporeal circuit. The damage to the polycarbonate connector had occurred after spillage of isoflurane during the filling of the vaporizer, causing it to crack and leak. The incident was managed by replacement of the cracked connector during a temporary circulatory arrest. The patient was hypothermic and the time off bypass was less than 1.5 min. There were no neurologic sequelae, the patient made an uneventful recovery. In conclusion, the spillage of volatile anesthetics can cause cracks in the polycarbonate connector of the extra-corporeal circuit, leading to potentially interruption of CPB.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Use of Volatile Anesthetic Agent in Extracorporeal Circuit as a Cause of Break in Polycarbonate Connector-Lessons Learnt.

Mishaps, near misses, and lethal incidents are known to occur during cardiopulmonary bypass. We share one such rare case of break in polycarbonate connector because of the use of isoflurane in extracorporeal circuit and its successful management.

متن کامل

SMART tubing presents an increased risk of disconnection during extracorporeal circulation.

A number of products exhibiting biocompatible features have been developed for use in extracorporeal blood circuits during cardiopulmonary bypass procedures. While attention has been focused on biocompatibility features of the blood-circuit interface, a number of issues applicable in clinical use of these circuits have arisen. Surface Modifying Additive Technology (SMART; Cobe Cardiovascular, A...

متن کامل

Oxygenator safety evaluation: a focus on connection grip strength and arterial temperature measurement accuracy.

This report describes the assessment of three specific safety-related specifications in the consideration of an alternate oxygenator; first the grip strength relationship between various oxygenator connectors and SMARxT tubing, second, the grip strength of various biopassive tubings and an isolated SMARxT connector, and finally, the accuracy of the arterial outlet temperature measurement. Grip ...

متن کامل

Unexplained Obstruction of an Integrated Cardiotomy Filter During Cardiopulmonary Bypass.

Cardiopulmonary bypass (CPB) is considered relatively safe in most cases, yet is not complication free. We present a case of an integrated cardiotomy filter obstruction during CPB, requiring circuit reconfiguration. Approximately an hour after uneventful initiation of CPB the integrated cardiotomy filter became obstructed over several minutes, requiring circuit reconfiguration using an external...

متن کامل

Ecmo in a diabetic patient with cerebellar infarction and ARDS- A case report

Background: Extra corporeal membrane oxygenation (ECMO) has been considered as a crucial modality for the management of acute respiratory distress syndrome (ARDS). Since ECMO is considered to be an invasive modality and may be associated with some complications, there are strict indications for its use. Here we report successful treatment of a case of severe ARDS with Veno-Venous ECMO. Case pre...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:

دوره 58  شماره 

صفحات  -

تاریخ انتشار 2010