A comparison of the dose distributions between the brachytherapy 125I source models, STM1251 and Oncoseed 6711, in a geometry lacking radiation equilibrium scatter conditions

نویسندگان

  • Kenichi Tanaka
  • Ken-ichi Kamo
  • Kunihiko Tateoka
  • Osamu Asanuma
  • Kaori Sato
  • Hiromitsu Takeda
  • Koh-ichi Sakata
  • Jun Takada
چکیده

The purpose of this study was to estimate the uncertainty in the dose distribution for the (125)I source STM1251, as measured with a radiophotoluminescent glass rod dosimeter and calculated using the Monte Carlo code EGS5 in geometry that included the source structure reported by Kirov et al. This was performed at a range of positions in and on a water phantom 18 cm in diameter and 16 cm in length. Some dosimetry positions were so close to the surface that the backscatter margin was insufficient for photons. Consequently, the combined standard uncertainty (CSU) at the coverage factor k of 1 was 11.0-11.2% for the measurement and 1.8-3.6% for the calculation. The calculation successfully reproduced the measured dose distribution within 13%, with CSU at k ≤ 1.6 (P > 0.3). Dose distributions were then compared with those for the (125)I source Oncoseed 6711. Our results supported the American Association of Physicists in Medicine Task Group No. 43 Updated Protocol (TG43U1) formalism, in which STM1251 dose distributions were more penetrating than those of Oncoseed 6711. This trend was also observed in the region near the phantom surface lacking the equilibrium radiation scatter conditions. In this region, the difference between the TG43U1 formalism and the measurement and calculation performed in the present study was not significant (P > 0.3) for either of the source models. Selection of the source model based on the treatment plans according to the TG43U1 formalism will be practical.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

محاسبات دزیمتری و مقایسه توزیع دز در اطراف دو مدل چشمه ید 125 در براکی تراپی

Brachytherapy is a therapeutic approach in which radioactive sources are used to irradiate malignant tumors from near distance. In this study, the comparison of dose distribution in iodine-125 source, STM 1251 and Oncoseed 6711 models was performed using the Monte Carlo Geant4 code. The results obtained from the Monte Carlo calculations have been compared with the TG43U1 formulation and also wi...

متن کامل

Determination of Dosimetric Characteristics of IrSeed 125I Brachytherapy Source

Introduction Low dose rate brachytherapy sources have been widely used for interstitial implants in tumor sites, particularly in prostate. Dosimetric characteristics of a new IrSeed 125I brachytherapy source have been determined using the LiF thermoluminescent dosimeter (TLD) chips. Materials and Methods Dose rate constant, radial dose function, and anisotropy function around the IrSeed 125I so...

متن کامل

New 125I brachytherapy source IsoSeed I25.S17plus: Monte Carlo dosimetry simulation and comparison to sources of similar design

PURPOSE To determine the relative dose rate distribution around the new (125)I brachytherapy source IsoSeed I25.S17plus and report results in a form suitable for clinical use. Results for the new source are also compared to corresponding results for other commercially available (125)I sources of similar design. MATERIAL AND METHODS Monte Carlo simulations were performed using the MCNP5 v.1.6 ...

متن کامل

Estimation of dosimetric parameters of I-125 brachytherapy source model 6711 using GATE8.1 code

Brachytherapy is one type of internal radiation therapy where radiation sources, which are usually encapsulated are placed as close as possible to the tumor site inside the patient's body. In this technique, it is important to determine dose distribution around the brachytherapy capsule. Hereby, in this paper, dosimetric parameters of I-125 brachytherapy source model 6711 are estimated accordin...

متن کامل

A dosimetry method for low dose rate brachytherapy by EGS5 combined with regression to reflect source strength shortage

The post-implantation dosimetry for brachytherapy using Monte Carlo calculation by EGS5 code combined with the source strength regression was investigated with respect to its validity. In this method, the source strength for the EGS5 calculation was adjusted with the regression, so that the calculation would reproduce the dose monitored with the glass rod dosimeters (GRDs) on a water phantom. T...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره 56  شماره 

صفحات  -

تاریخ انتشار 2015