Can we re-design the gradient coil to make the eddy current field match the primary gradient field?

نویسندگان

  • H. Sanchez
  • M. Poole
  • A. Trakic
  • S. Crozier
چکیده

Introduction: Rapidly switched magnetic field gradients induce eddy currents in the nearby conducting structures of the MRI scanner. These eddy currents in turn generate an opposing and distorting transient magnetic field in the region of interest (ROI) with consequential image distortion. A tailored current pulse is usually applied to minimize the undesired effects of the eddy currents in the ROI. In order to successfully apply this current compensation technique, the primary and the secondary magnetic field produced by the eddy currents should have a similar spatial form in the ROI (field matching). Some efforts have been made to improve the linearity of the secondary and hence maximize the field matching figure [1,2]. In this work we present two approaches for gradient coil design that produces gradient fields with characteristics similar to those produced by the eddy currents. The first approach iteratively refines the gradient coil so that its primary field is similar to the induced secondary field. The second approach, termed as the mode coil [3], classifies as an eigenvector assignment [4] technique. The mode coil [3] produces maximal field matching with a large field non-linearity. This limitation is improved by designing an actively shielded mode gradient coil. The presented examples are used to demonstrate the advantages and disadvantages of the proposed techniques. Method: It is assumed that a 15 mm thick aluminum cylinder (surface C) of radius 377.5 mm and half length of 550 mm surrounds a cylindrical coil surface S of radius 320 mm and half length of 500 mm. When a low-frequency driving current is applied to the gradient set, the skin depth is very small under the aforementioned assumption, and

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Eigenmode Analysis of Eddy Currents and Eigenmode Coil Design

Introduction Eddy currents have long been the bane of MRI. Particularly since the mid 1980s when imaging sequences became more rapid [1] and superconducting magnets became routinely used. When gradient coils are switched rapidly, eddy currents are induced in the cryostat vessel and other surrounding conducting structures. Two principal solutions have been adopted into standard practice for deal...

متن کامل

Discrete Design Method of a Transverse Gradient Coil for Magnetic Resonance Imaging

S. Shvartsman, G. DeMeester, M. Morich Philips Medical Systems, Cleveland, Ohio, United States Introduction In this work we describe a discrete gradient coil design method that does not include some of the essential steps of continuous current based minimum energy design methods. The key concept is that all characteristics of a transverse gradient, such as coil inductance, spatial distribution ...

متن کامل

Adaptive Method for Gradient Coil Design

Method One of the obstacles in designing gradient coil is the general discretization procedure that approximates the continuous current densities by a number of current carrying conductors. The shape of the conductors’ layout determines the properties of the gradient coil such as the gradient strength, field quality characteristics inside the FoV, level of shielding, coil resistance, slew rate,...

متن کامل

Temporal analysis of Z-Gradient coil eddy currents in tungsten collimator with different resistivities for SPECT/MRI

INTEC, Ghent University/iMinds, Ghent, Belgium Combining Single Photon Emission Computed Tomography (SPECT) with Magnetic Resonance Imaging (MRI) results in an interaction of the time-varying magnetic field gradients with the highly conducting tungsten collimator, which generates a secondary magnetic field causing spatial distortions in reconstructed MR images. Accurate simulations are importan...

متن کامل

Eddy-Current-Compensated RF Pulse Design for Parallel Excitation

Introduction: High-performance RF coils and shielded Gradient coils [1] for high or ultra high field MRI often require the use of RF shields that are in close proximity to the imaged volume. These shields can sometimes generate Eddy-Currents that are not adequately compensated for using the pre-emphasis algorithm [2] of the scanner. K-space trajectory measurements [3] can be used to compensate ...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:

دوره   شماره 

صفحات  -

تاریخ انتشار 2009