Assessment of potential jaw‐tracking advantage using control point sequences of VMAT planning
نویسندگان
چکیده
This study aims to evaluate the potential jaw-tracking advantage using control point sequences of volume volumetric modulated arc therapy (VMAT) planning. VMAT plans for patients with prostate and head and neck (H&N) cancers were converted into new static arc (SA) plans. The SA plan consisted of a series of static fields at each control point of the VMAT plan. All other machine parameters of the SA plan were perfectly identical to those of the original VMAT plan. The jaw-tracking static arc (JTSA) plans were generated with fields that closed the jaws of each SA field into the multileaf collimators (MLCs) aperture. The dosimetric advantages of JTSA over SA were evaluated in terms of a dose-volume histogram (DVH) of organ at risk (OAR) after renormalizing both plans to make the same target coverage. Both plans were delivered to the MatriXX-based COMPASS system for 3D volume dose verification. The average jaw size reduction of the JTSA along the X direction was 3.1 ± 0.9 cm for prostate patients and 6.9 ± 1.9 cm for H&N patients. For prostate patients, the organs far from the target showed larger sparing (3.7%-8.1% on average) in JTSA than the organs adjacent to the target (1.1%-1.5%). For the H&N plans, the mean dose reductions for all organs ranged from 4.3% to 11.9%. The dose reductions were more significant in the dose regions of D80, D90, and D95 than the dose regions of D5, D10, and D20 for all patients. Likewise, the deliverability and reproducibility of jaw-tracking plan were validated. The measured dosimetric advantage of JTSA over SA coincided with the calculated one above.
منابع مشابه
A comparative study of identical VMAT plans with and without jaw tracking technique
The unwanted radiation transmission through the multileaf collimators could be reduced by the jaw tracking technique which is commercially available on Varian TrueBeam accelerators. On the basis of identical plans, this study aims to investigate the dosimetric impact of jaw tracking on the volumetric-modulated arc therapy (VMAT) plans. Using Eclipse treatment planning system (TPS), 40 jaw-track...
متن کامل4D VMAT planning and verification technique for dynamic tracking using a direct aperture deformation (DAD) method
We developed a four-dimensional volumetric modulated arc therapy (4D VMAT) planning technique for moving targets using a direct aperture deformation (DAD) method and investigated its feasibility for clinical use. A 3D VMAT plan was generated on a reference phase of a 4D CT dataset. The plan was composed of a set of control points including the beam angle, MLC apertures and weights. To generate ...
متن کاملDosimetric Effect of Jaw Tracking in Volumetric-Modulated Arc Therapy
The aim of this study was to investigate the potential of jaw tracking with the volumetric-modulated arc therapy (VMAT) to reduce the normal tissue dose. Plans of nasopharynx, lung, and prostate cancers (10 plans for each) were used to perform VMAT with and without jaw tracking. The dose reduction was evaluated in terms of organ doses and integral doses. Organ-dose reduction with jaw tracking w...
متن کاملDosimetric Evaluation of Volumetric Modulated Arc Therapy (VMAT) and Intensity Modulated Radiotherapy (IMRT) Using AAPM TG 119 Protocol
Background: The commissioning accuracy of Volumetric Modulated Arc Therapy (VMAT) need to be evaluated.Objective: To test and evaluate commissioning accuracy of VMAT based on the TG 119 protocols at local institution. Material and Methods: The phantom, structure sets, VMAT and IMRT beam parameter setup, dose prescriptions and planning objectives were following TG 119 guidelines to c...
متن کاملInfluence of jaw tracking in intensity-modulated and volumetric-modulated arc radiotherapy for head and neck cancers: a dosimetric study
PURPOSE To Study the dosimetric advantage of the Jaw tracking technique in intensity-modulated radiotherapy (IMRT) and volumetric-modulated arc therapy (VMAT) for Head and Neck Cancers. MATERIALS AND METHODS We retrospectively selected 10 previously treated head and neck cancer patients stage (T1/T2, N1, M0) in this study. All the patients were planned for IMRT and VMAT with simultaneous inte...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
دوره 15 شماره
صفحات -
تاریخ انتشار 2014