Micromotion of cementless tibial baseplates under physiological loading conditions.
نویسندگان
چکیده
Initial implant stability is crucial to cementless knee arthroplasty success. The objective of this study was to develop a physiological relevant methodology that incorporates torsion, shear, and compression forces to evaluate two tibial component designs that feature either a keel or cylindrical porous metal pegs. The data were compared with a simplified compression loading scenario. Results show a loading profile that combines compressive, shear, and torsional loads results in significantly larger motions than occur when loading in compression only. When comparing between a keeled and a pegged device, the new method shows significant differences in tibial component subsidence/liftoff at the anterior and posterior locations, which were lacking in the simplified test model. To accurately assess implant stability, studies should use physiological relevant loading.
منابع مشابه
•The Effect of Peg vs. Keel Fixation on Micromotion of Cementless Tibial Baseplates
INTRODUCTION: Initial implant stability is crucial to cementless knee replacement success, however the optimal design remains elusive. Previous biomechanical studies have aimed at quantifying initial stability using simplified loading scenarios. The objective of this study was to use a physiological relevant methodology which incorporates torsion, shear and compression forces to evaluate tibial...
متن کاملNovel Method for Testing Cementless Tibial Components Using Micromotion and Impedance Vibration Analysis
INTRODUCTION: The initial mechanical stability of cementless implants is critical to achieve successful osseointegration through minimization of deleterious micromotion. Controversy exists regarding the most reliable and appropriate methodology for in vitro mechanical testing of implants used to guide design and performance of cementless total knee replacement. Mechanical testing methods vary w...
متن کاملEffect of Tibial Tray Malalignment on Micromotion and Subsidence: A Preliminary Report
Malalignment of the tibial tray in total knee arthroplasty has been linked to several complications including tibial tray subsidence, leading to revision surgery. However, quantitative biomechanical evidence to directly support the mechanism of failure is not available. We developed a model to study tibial tray micromotion and subsidence in vitro under multiaxial physiological loading condition...
متن کاملDevelopment of Six Degree of Freedom Mechanical Testing for Assessing The Primary Stability of Cementless Tibial Trays
Introduction: Cementless tibial fixation has been used for over 30 years in total knee arthroplasty. There are several potential advantages including preservation of bone stock and ease of revision. More importantly, for young active patients there is the potential for increased longevity of fixation. However, the clinical results have been variable, with reports of extensive radiolucent lines,...
متن کاملPatient-Specific Unicompartmental Knee Resurfacing Arthroplasty: Use of a Novel Interference Lock to Reduce Tibial Insert Micromotion and Backside Wear
Micromotion has long been associated with wear of polyethylene tibial inserts, potentially causing failure of unicompartmental knee replacement systems. One cause of micromotion is the locking mechanism between the undersurface of the polyethylene and the tibial tray. The objective of this study was to investigate the use of new novel lock designs for reducing the micromotion associated with a ...