PET/MR attenuation correction in brain imaging using a continuous bone signal derived from UTE
نویسندگان
چکیده
Department of Clinical Physiology, Nuclear Medicine and PET, Rigshospitalet Copenhagen, Copenhagen, Denmark In the absence of transmission sources in combined clinical PET/MR systems, MR images are used for MR-based attenuation correction (MRAC). The main challenge in MR-AC is to separate the bone and air, as neither have a signal in the MR images. In the attenuation maps supplied by the vendor, a single value is assigned to bone using an ultra-short echo time (UTE) MR sequence. The purpose of this study was to develop a new multi-class segmentation-based MR-AC method, employing Continuous-Bone-using-R2* (MRAC_CBuR2*), and evaluate it on a large patient cohort. 53 [18F]-FDG PET/MR brain patients were included in this study. MRAC was based on an aligned CT (MRAC_CT, used as reference), standard MRAC_UTE and MRAC_CBuR2*. Our method segments the air, brain, CSF and soft tissue voxels on the UTE images, and uses a mapping of R2* values to HU to measure the density in bone voxels. Aligned anatomical masks are used to improve accuracy in noisy regions. Region-based analysis was performed using ICBM 2009a brain atlas with anatomical labels pre-defined. Using CBuR2*, 82% of the voxels in the brain are within ±5% of PET_CT, compared to 27% when using UTE. Using our method, there are clear improvements over UTE. The average error over the full brain is 0.8% (±1.7%), compared to -7.1% (±2.4%) in UTE. Of note, the maximum error in the cerebellum is -15% and 7% in UTE and CBuR2*, respectively. The proposed method uses the available UTE images to segment tissue classes, and uses the R2* map to measure a continuous bone signal. The improvement over the vendor provided UTE reduces both the global and local error on the reconstructed PET images.
منابع مشابه
Comparison of state-of-the-art atlas-based bone segmentation approaches from brain MR images for MR-only radiation planning and PET/MR attenuation correction
Introduction: Magnetic Resonance (MR) imaging has emerged as a valuable tool in radiation treatment (RT) planning as well as Positron Emission Tomography (PET) imaging owing to its superior soft-tissue contrast. Due to the fact that there is no direct transformation from voxel intensity in MR images into electron density, itchr('39')s crucial to generate a pseudo-CT (Computed Tomography) image ...
متن کاملRegion specific optimization of continuous linear attenuation coefficients based on UTE (RESOLUTE): application to PET/MR brain imaging
The reconstruction of PET brain data in a PET/MR hybrid scanner is challenging in the absence of transmission sources, where MR images are used for MR-based attenuation correction (MR-AC). The main challenge of MR-AC is to separate bone and air, as neither have a signal in traditional MR images, and to assign the correct linear attenuation coefficient to bone. The ultra-short echo time (UTE) MR...
متن کاملRegion specific optimization of continuous linear attenuation coefficients based on UTE (RESOLUTE): application to PET/MR brain imaging.
The reconstruction of PET brain data in a PET/MR hybrid scanner is challenging in the absence of transmission sources, where MR images are used for MR-based attenuation correction (MR-AC). The main challenge of MR-AC is to separate bone and air, as neither have a signal in traditional MR images, and to assign the correct linear attenuation coefficient to bone. The ultra-short echo time (UTE) MR...
متن کاملMR-based attenuation correction in brain PET based on UTE sequences
Department of Nuclear Medicine, Klinikum rechts der Isar, Technische Universität München, Germany Attenuation correction (AC) in brain PET/MR has recently emerged as one of the challenging tasks in the PET/MR field. It has been shown that to ignore the attenuation produced by bone can lead to errors ranging from 5-30% in regions close to bone structures. Since the information provided by the MR...
متن کاملMR-based PET Attenuation Correction for Neurological Studies Using Dual-Echo UTE Sequences
INTRODUCTION Due to the limited space available inside an MR scanner, most of the MR compatible PET inserts are not equipped with a transmission source, which makes the implementation and validation of an MR-based attenuation correction (AC) method necessary. The obvious challenge is that MR images are not typically directly related to tissue linear attenuation coefficients (LACs). Furthermore,...