Evaluation of MicroScan rapid gram-positive panels for detection of oxacillin-resistant staphylococci.

نویسندگان

  • G L Woods
  • D LaTemple
  • C Cruz
چکیده

The ability of MicroScan rapid panels (Baxter MicroScan, West Sacramento, Calif.) to detect oxacillin resistance in 92 clinical isolates of Staphylococcus aureus and 103 coagulase-negative staphylococci was evaluated by comparing results with those of MicroScan 24-h MIC panels and, to resolve discrepancies, oxacillin agar screening. Both panels were interpreted by the MicroScan WalkAway-96 system. Rapid panels detected 96.7% of resistant S. aureus isolates and 72% of resistant coagulase-negative staphylococci, 22 of which did not grow in the panels.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Performance of eight methods, including two new rapid methods, for detection of oxacillin resistance in a challenge set of Staphylococcus aureus organisms.

Using a set of 55 Staphylococcus aureus challenge organisms, we evaluated six routine methods (broth microdilution, disk diffusion, oxacillin agar screen, MicroScan conventional panels, MicroScan rapid panels, and Vitek cards) currently used in many clinical laboratories and two new rapid methods, Velogene and the MRSA-Screen, that require less than a day to determine the susceptibility of S. a...

متن کامل

Clinical importance of identifying coagulase-negative staphylococci isolated from blood cultures: evaluation of MicroScan Rapid and Dried Overnight Gram-Positive panels versus a conventional reference method.

We evaluated the clinical usefulness of species identification of blood isolates of coagulase-negative staphylococci as a predictor of the clinical significance of the isolates. In addition, we compared results of species identification obtained with MicroScan Rapid Gram-Positive Identification panels and Dried Overnight (Conventional) Gram-Positive Identification panels with those obtained by ...

متن کامل

Evaluation of the E test by using selected gram-positive bacteria.

The E test (AB Biodisk NA Inc.) was compared with standard reference methods using the National Committee for Clinical Laboratory Standards's recommendations for determining the MICs of four selected antibiotics against 208 clinical isolates of gram-positive bacteria. These bacteria included 32 strains of Streptococcus pneumoniae, 25 strains of Enterococcus faecium, 20 strains of oxacillin-sens...

متن کامل

Early Screening of oxacillin-resistant Staphylococcus aureus and Staphylococcus epidermidis from blood culture.

The timely detection of blood-borne pathogens is one of the most important functions of the microbiology laboratory. Recently, methicillin-resistant staphylococci have become the most important pathogens seen by the laboratory. The purpose of this study was to evaluate Staphy agar, a novel screening medium, for the detection methicillin-resistant Staphylococcus aureus, S. epidermidis, or other ...

متن کامل

Evaluation of the AutoMicrobic system Gram-Positive Susceptibility-MIC card for detection of oxacillin-resistant coagulase-negative staphylococci.

A total of 398 consecutive clinical staphylococcus isolates, of which 205 were coagulase negative and 193 were coagulase positive, were tested in parallel by using AutoMicrobic system Gram-Positive Susceptibility-MIC cards and modified Mueller-Hinton agar containing 4% NaCl and oxacillin (6 micrograms/ml). The AutoMicrobic system cards correctly detected 103 of 104 (99%) oxacillin-resistant coa...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Journal of clinical microbiology

دوره 32 4  شماره 

صفحات  -

تاریخ انتشار 1994