In Vivo Quantification of Renal Sodium Concentration with a Dual Rf Resonator System

نویسندگان

  • R. Kalayciyan
  • F. Wetterling
  • S. Neudecker
  • L. R. Schad
چکیده

Introduction: The quantification of local Tissue Sodium Concentration (TSC) levels can serve as a diagnostic measure for renal diseases, because TSC is strongly linked to the corticomedullary function, which regulates the vascular ion concentrations [1, 2]. However, the validation of the achieved TSC quantification accuracy on rat kidneys using quantitative Na Magnetic Resonance Imaging (qNa-MRI) technique is challenging and results in low SNR in acquired Na images. Additionally, homogeneous B1-field is required for quantification of the TSC. Although surface coils provide high SNR, the generated B1-field of a surface coil is highly inhomogeneous. Therefore, the aim of this work was to develop a dual RF resonator system composed of a noise-matched receive-only surface coil and a commercial double-tuned H/Na transmit volume resonator in order to maximize signal sensitivity as well as the B1-field homogeneity for TSC measurements in in vivo rat kidneys. Methods: Noise-matching via low noise preamplifiers (LNPs) introduced by Roemer et al. is a commonly used concept for decoupling multiple receiver elements in phased array coils [3]. In this work, the developed Na receive-only surface coil was designed to be a single-coil element (rectangular 25mm x 50mm). The LNP (MwT, USA, 2.1 Ω input impedance, 29 dB gain, 0.5 noise figure) was noise-matched to the detector element via a λ/4 transformation network, while the by Reykowski et al. [4] suggested pi-transformation-network was modified with the intention to minimize the number of electronic circuit components. Thus, the loop-coil was noise-matched to the required 50 Ω input impedance of the preamplifier, which when connected to the transformation network presents high impedance to the coil at its terminals. During the design the transformation path of the network was adjusted to the resonance frequency of 105 MHz. The efficiency of this trap circuit was measured by a S21 measurement with decoupled probe pair (4 cm Ø) brought to near proximity of the receive-only coil: The peak split was measured to be 20 MHz and -45 dB preamplifier decoupling was achieved. Furthermore, passive decoupling circuit was added to the coil circuit to compensate for imperfect geometric decoupling of the surface coil from the volume resonator. The birdcage resonator’s B1-field was orthogonally arranged to the surface coil’s B1-field in order to maximize the geometric decoupling. The Na-imaging performance of the LNP coil was compared to a home-built transceiver surface coil (Qloaded / unloaded = 100 / 112) with the identical detector dimensions. A cylindrical tube of 4 cm Ø filled with 155 mMol/l NaCl solution was scanned with 3D-GRE sequence on a 9.4 T system (Bruker BioSpin GmbH, Ettlingen, Germany) with TE/TR = 3ms/20ms, voxel size = 1 x 1 x 2.6 mm, and an acquisition time of 10 min (Fig. 1a-c). To provide a comparable situation for both coil system sensitivities, the phantom experiments were performed in existence of the volume resonator for both transceiver (TxRx) and transmit-only-receive-only (ToRo) measurements. A further measurement was performed using the double-tuned Na/H transceiver volume resonator (Bruker BioSpin GmbH, Ettlingen, Germany) with inner diameter 72mm in order to estimate its B1-field homogeneity (see Fig. 1c). The flip angle calibration for the ToRo and volume transceiver coil experiments was straight forward, but for the transceiver surface coil inhomogeneous B1-field hampered exact flip angle calibration so that the 90° flip angle had to be set in approximately 4mm distance from the surface coil’s detector element. The SNR values were plotted as a function of depth along the vertical profile through the centre of the object and are graphed in Fig. 1d. In order to quantify the TSC the surface coil’s receive sensitivity was compensated for through a reference scan method [5]. A homogeneous reference phantom was scanned in addition to the two fiducial vials (155 mMol NaCl), which were permanently fixed on top of the receiver coil (see Fig. 2b, d). Sample and reference scans were co-registered to the reference vial positions before the sensitivity corrected Na image was computed as the quotient image (Fig. 2c) from the sample and the reference scans, respectively. Differences in coil loading between both sample and reference scan were compensated for by a correction factor [6]. To demonstrate the proof-of-concept, TSC quantification accuracy was verified for imaging of a concentration phantom containing gels of various NaCl and agarose concentrations (40-160 mM NaCl in 0-5% agarose gel) shown in Fig. 2a. H-reference scan was recorded with Na/H transceiver volume resonator by means of a 3D-Flash sequence using TE/TR = 3 / 20 ms, anisotropic resolution of 0.5x0.5x1mm, and Tacq = 3 min. 3D-UTE sequence which is an FID readout technique sampling the k-space in a radial manner starting from the k-space center [7] was used for in vivo and phantom Na MRI acquisitions. The radial 3D ultrashort TE (3D-UTE) sequence is less prone to motion artifacts than the 3D-Gradient Echo (GRE) sequence, and the fact that 3D-UTE allowed for short TE < 0.1 ms. A healthy female rat (~350 g) was imaged in vivo with following sequence parameters: isotropic 1mm voxel size, TR = 60 ms, Tacq = 13 min, BW = 5 kHz, 0.1 ms block pulse length, FOV = 6.4 cm and TEUTE = 60μs. Results and Discussion: The dual RF resonator system for Na MRI comprised a modified concept for noise-matching the Na receive-only surface coil to the LNP. In conjunction with the volume resonator, optimum B1-field homogeneity was generated, while high SNR Na images could be acquired an essential aspect for the quantification of TSC in the rat kidney. The SNR of the TxRx surface coil was measured to be approx. 15% lower compared to the ToRo system which may be the result of undesired coupling with the volume resonator that was left inside the magnet during the TxRx measurement (see Fig. 1d). Furthermore, the ToRo system achieved slightly better penetration into depth of the sample which is a consequence of the provided homogeneous B1-field by the volume resonator in this mode. The quantitative measurements of known phantom Na concentrations resulted in reasonably good quantification precision of ±10 %. After verification of the quantification technique on a phantom, in vivo TSC was quantified for a healthy rat (see Fig. 3 right). The corticomedullary TSC gradient with 90 mM TSC maximum is well-recognizable in the TSC map of the rat kidney (Fig. 3c). Using short TE (60 μs) in combination with relatively long TR (60 ms) enabled the TSC quantification without the need to compensate for T1 or T2 relaxation effects. However, the quantification method, especially the co-registration of the reference and sample images, is prone to errors caused by the close positioning of the fiducial vials to the resonator and consequently the variation in the B1-field in these regions. Future quantification studies need to include a third vial with known Na concentration within the field of view to improve the TSC quantification accuracy. References: [1] Maril et al., Kidney Int. 69:765-768 (2006); [2] Maril et al., Kidney Int. 65:927-935 (2004); [3] Roemer et al., MRM, 16(2):192-225 (1990); [4] Reykowski et al., MRM, 33(6):848-852 (1995); [5] Wetterling et al., Proc. ISMRM 18 (2010); [6] Ouwerkerk et al., Breast Cancer Res Treat. 106 (2007); [7] Rahmer et al., MRM 55:1075-1082 (2006).

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

A double-tuned (1)H/(23)Na dual resonator system for tissue sodium concentration measurements in the rat brain via Na-MRI.

A method for quantifying the tissue sodium concentration (TSC) in the rat brain from ²³Na-MR images was developed. TSC is known to change in a variety of common human diseases and holds considerable potential to contribute to their study; however, its accurate measurement in small laboratory animals has been hindered by the extremely low signal to noise ratio (SNR) in ²³Na images. To address th...

متن کامل

Multipactor Analysis in Dielectric Resonator Waveguide Filters

Multipactor breakdown analysis is presented in cylindrical dielectric resonator waveguide filter under its dual-mode behavior. The method of effective electron algorithm with combination of Monte-Carlo method is used to perform the simulation and predict multipactor breakdown thresholds. The breakdown thresholds of the proposed structure are presented for different values of frequency-gap produ...

متن کامل

In Vivo Sodium Imaging of Kidney Using 3d Ultrshort Echo Time Sequence

Introduction: Local Tissue Sodium Concentration (TSC) levels could serve as a biomarker for tissue viability and integrity in stroke [1] and tumor [2]. It is therefore anticipated that TSC measurements can also help to diagnose renal diseases, since vascular sodium concentration is regulated by the corticomedullary sodium concentration gradient [3,4]. However, the SNR in Na Magnetic Resonance I...

متن کامل

Fractal Bandpass Filter Using Y-shaped Dual-Mode Resonator for C-Band Receiver (Research Note)

In this study, a fractal, Y-shaped dual-mode resonator bandpass filter (BPF) with input-output cross-coupling is introduced. A parallel-coupling feed structure with a cross coupling has been used to generate two transmission zeroes (TZs) near the lower and upper cutoff frequency that can effectively improve the passband edge selectivity. Also, a fractal shaped based on conventional diamond and ...

متن کامل

Performance of a Dual Chamber Microbial Fuel Cell using Sodium Chloride as Catholyte

Microbial fuel cell represents an emerging technology to attain electrical energy from wastewater. There are several alternative methods available for wastewater treatment; Microbial fuel cell is one of them, which generates green energy from wastewater for making a contribution to renewable sources of energy. This study states the performance of microbial fuel cell with different parameters i....

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:

دوره   شماره 

صفحات  -

تاریخ انتشار 2010