Evaluation of accuracy and reproducibility of a relocatable maxillary fixation system for fractionated intracranial stereotactic radiation therapy
نویسندگان
چکیده
INTRODUCTION Accurate localisation is an essential component for the delivery of intracranial stereotactic treatment. For fractionated stereotactic radiotherapy, we compared the daily localisation accuracy of a standard thermoplastic mask with a new maxillary fixation device (MFD). METHODS Daily pre-treatment kV cone-beam computed tomography (CBCT) scans of 23 patients (12 localised in the MFD and 11 in the mask) with benign skull-based lesions were reviewed retrospectively. The set up accuracy was measured in 6° of freedom, to ascertain both individual and population random and systematic errors. The appropriate clinical target volume to planning target volume margin was computed from set up error data. RESULTS A total of 682 CBCT scans were evaluated. Systematic (Σ) and random (σ) population errors were Σ = 0.8 mm, 0.2 mm and 0.2 mm and σ = 0.3 mm, 0.3 mm and 0.2 mm, respectively, for the standard mask in the left/right (LR), superior/inferior (SI), and anterior/posterior (AP) translational planes, and Σ = 0.2 mm, 0.1 mm and 0.2 mm and σ = 0.2 mm, 0.3 mm and 0.2 mm, respectively, for the MFD. There was a reduction in rotation errors in the MFD compared to the mask. Margin calculations suggested an isotropic margin could be safely reduced to 2 mm for the MFD. CONCLUSION The two devices demonstrate similar daily positional accuracy for fractionated stereotactic treatment of intracranial lesions. Combined with daily image guidance and couch correction, either of these devices is a viable frameless option for fractionated stereotactic radiation therapy.
منابع مشابه
Stereotactic Radiosurgery/Radiotherapy: A Historical Review
"Stereotactic" is an exact radiotherapy treatment modality which implements invasive and non-invasive facilities for improving precise dose delivery. Stereotactic refers to three-dimensional localization of a specific point in space by a unique set of coordinates that relate to a fixed external reference frame. An accurate delivery of radiation is attainable using these techniques with high pre...
متن کاملReproducibility and geometric accuracy of the fixster system during hypofractionated stereotactic radiotherapy
BACKGROUND Hypofractionated radiotherapy has been used for the treatment of AVMs and brain metastases. Hypofractionation necessitates the use of a relocatable stereotactic frame that has to be applied on several occasions. The stereotactic frame needs to have a high degree of reproducibility, and patient positioning is crucial to achieve a high accuracy of the treatment. METHODS In this study...
متن کاملStereotactic radiotherapy in the treatment of ocular melanoma: A noninvasive eye fixation aid and tracking system
Ocular melanoma is frequently treated using brachytherapy implants (such as 125I and 60Co plaques or 184Ta wire), surgery, or external beam radiotherapy using small 60Co beams, high energy x-rays, or proton therapy. The last technique, though very expensive, provides improved dose distributions and dose localizations in the treatment of tumours adjacent to critical normal tissues. The technique...
متن کاملClinical Positioning Accuracy for Multisession Stereotactic Radiotherapy With the Gamma Knife Perfexion
Multisession stereotactic radiation therapy is increasingly being seen as a preferred option for intracranial diseases in close proximity to critical structures and for larger target volumes. The objective of this study is to investigate the reproducibility of the Extend system from Elekta. A retrospective review was conducted for all patients treated with multisession Gamma Knife between July ...
متن کاملDevelopment of A Head Docking Device for Linac-based Stereotactic Radiosurgery System to be Installed on Neptun 10 PC Linac
Introduction: Stereotactic radiosurgery is a technique for the treatment of intracranial lesions with highly collimated x-ray beam. Linac-based radiosurgery is currently performed by two techniques: couch or pedestal mounted. In the first technique, a device is required to affix patient's head to a couch and moreover to translate it accurately. Structure of suc...