BACKGROUND The EasyTube® (EzT) is a supraglottic airway device (SAD) enabling ventilation irrespective of its placement into the esophagus or trachea. Data obtained on SADs from multicenter studies, performed in highly specialized centers cannot always be transferred to other sites. However, data on comparability of different sites are scarce. This study focused on inter-site variability of ventilatory and safety parameters during general anesthesia with the EzT. METHODS 400 patients with ASA physical status I-II undergoing general anesthesia for elective surgery in four medical centers (EzT group (n = 200), ETT group (n = 200)). Mallampati classification, success of insertion, insertion time, duration of ventilation, number of insertion attempts, ease of insertion, tidal volumes, leakage, hemodynamic parameters, oxygenation, and complications rates with the EasyTube (EzT) or endotracheal tube (ETT) in comparison within the sites and in between the sites were recorded. RESULTS Intra-site and inter-site comparison of insertion success as primary outcome did not differ significantly. The inter-site comparison of expiratory minute volumes showed that the volumes achieved over the course of anesthesia did not differ significantly, however, mean leakage at one site was significantly higher with the EzT (0.63 l/min, p = 0.02). No significant inter-site differences in heart rate, blood pressure, or oxygenation were observed. Sore throat and blood on the cuff after removal of the device were the most frequent complications with significantly more complications at one site with the EzT (p = 0.01) where insertion was also reported significantly more difficult (p = 0.02). CONCLUSION Performance of the EzT but not the ETT varied between sites with regard to insertion difficulty, leakage, and complications but not insertion success, ventilation, hemodynamics, and oxygenation parameters in patients with ASA physical status 1-2 during general anesthesia undergoing minor elective surgery.
منابع مشابه
The EasyTube during general anesthesia for minor surgery
BACKGROUND The EasyTube (EzT) is a supraglottic airway device that is used for emergency airway situations. Ventilation during general anesthesia should also be feasible, but literature on the EzT is scarce. We evaluated the EzT in comparison with the endotracheal tube (ETT) in its use during general anesthesia in a comparative study. METHODS A total of 400 patients with American Society of A...
متن کاملTapia's Syndrome after Corrective Jaw Surgery under General Anesthesia: A Case Report
Introduction:Tapia’s syndrome is a rare complication of recurrent laryngeal and hypoglossal nerve paralysis due to anesthetic airway mismanagement or malpositioning of the patient’s head during surgery. Case Report:Here we present a case of Tapia's syndrome in a 22-year-old male after corrective jaw surgery under general anesthesia, with a long period of recovery, related to airway management p...
متن کاملمقایسه اثر استفاده از ماسک حنجره (Laryngeal mask airway) و لوله کافدار حلقی (Cuffed pharyngeal tube) در بروز گلودرد پس از بیهوشی عمومی در بزرگسالان
Post-operative sore throat is one of the most common complications and complaints of patients after general anesthesia especially in operations that need endotracheal intubations. Its causes are: size of endotracheal tube and type of its cuff, inadequate airway humidification, trauma during intubation and suctioning, high flow of inspiratory gases, surgical manipulation of airway and adjacent o...
متن کاملComparison of Effectiveness of Betamethasone gel Applied to the Tracheal Tube and IV Dexamethasone on Postoperative sore Throat: A Randomized Controlled Trial
Introduction: Postoperative sore throat is a common complaint in patients with endotracheal intubation and has potentially dangerous complications. This randomized controlled trial study investigated the incidence of postoperative sore throat after general anesthesia when betamethasone gel is applied to a tracheal tube compared with when IV dexamethasone is prescribed. Materials and Methods: ...
متن کاملNew Metal Shield for the Endotracheal Tube during Laser Endolaryngeal Surgery
Introduction: To present and assess a newly designed autoclavable, reusable tube shield for an endotracheal tube (ETT) during laser laryngeal surgery. Materials and Methods: This study included patients scheduled for endolaryngeal laser surgery. A carbonated stainless-steel hemisphere shield (~1.5×0.6 mm) connected to a silk thread passing through a hole in its middle near upper edge was ...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
دوره 12 شماره
صفحات -
تاریخ انتشار 2017