Endoscopic removal of a broken self-expandable metal stent using the stent-in-stent technique.
نویسندگان
چکیده
A 55-year-old man with dysphagia due to an unresectable distal esophageal adenocarcinoma underwent placement of a 15-cm partially covered self-expandable metal stent (SEMS), diameter 18mm (Ultraflex; Boston Scientific, Natick, Massachusetts, USA). After 3 months, he presented with recurrent dysphagia. At endoscopy, the meshes of the SEMS had broken, leading to collapse of the SEMS and obstruction of the lumen (●" Fig.1). Because the patient’s clinical condition had improved and further treatment with chemotherapy or radiotherapy was considered, we decided to remove the SEMS using the stent-in-stent technique. A second SEMS, 17-cm long, fully covered, diameter 20mm (Hanaro, M. I.Tech, Korea) was inserted, overlapping the full length of the first SEMS. After 2 weeks, on endoscopic examination, the loose struts of the disrupted SEMS had perforated the covering of the fully covered SEMS, and the radial pressure of the fully covered SEMS had succeeded in detaching the partially covered SEMS from the esophageal wall and tumor (●" Fig.2). By pulling the proximal retrieval lasso of the fully covered SEMS, both SEMSs were extracted simultaneously without the need to exert much force. Examination of the retrieved stents showed they were strongly attached to each other, indeed with loose struts of the first stent perforating the covering of the fully covered SEMS (●" Fig.3). Immediately after removal of both stents, no esophageal stenosis was observed. The post-procedural course was uneventful, and the patient is currently without dysphagia. Spontaneous fracture with collapse of an esophageal SEMS is a rare complication [1,2]. However, if a SEMS is extracted by pulling its upper or lower end, and traction is forceful, as in the case of (partial) stent ingrowth, the struts may break causing the metal wire mesh to become disrupted. When the struts are already broken, it seems likely that by pulling on one end of the SEMS, the metal framework will be destroyed even further, making it impossible to remove the SEMS all in one. In such cases, the stent-in-stent technique is a simple and safe method to facilitate removal of embedded stents in benign disease [3]. This is the first report of removal of a broken SEMS using the stent-in-stent technique. It appears to be simple, safe and effective, also in malignant disease.
منابع مشابه
Subclavian artery-esophageal fistula after placement of a self-expanding metal stent in a patient with esophagogastric anastomosis stenosis
Background: There have been reports on stent-related vascular erosions about patients with benign or malignant stenosis of the esophagus who received endoscopic stent insertion for palliative intention for oral intake. Case presentation: A 61-year-old woman with esophageal cancer located in the middle part of esophagus was treated with esophagectomy. Two years following the surgery, malignan...
متن کاملFracture of Self-Expandable Metal Stent during Endoscopic Removal in Benign Biliary Stricture
The endoscopic insertion of the self-expandable metal stent (SEMS) in benign biliary stricture has become an alternative to surgery. Fracture or migration of SEMS can occur rarely as complications. We report a case of fracture of SEMS during endoscopic retrieval in patients with chronic pancreatitis. In this case, broken stent was successfully removed with endoscopic ballooning of bile duct and...
متن کاملAcquired Infrarenal Abdominal Aortic Coarctation: Treatment with Percutaneous Self Expandable Stent
Abdominal aortic coarctation is an extremely rare vascular pathology. Its etiology can be congenital or acquired. Here we present a case of acquired infrarenal abdominal coarctation in a 66-year-old woman who complained of low back and legs pain. She had no signs of resting lower limb ischemia, with diminished distal pulses and normal blood pressure in upper and lower extremities. Magnetic reso...
متن کاملSafety of endoscopic removal of self-expandable stents after treatment of benign esophageal diseases.
BACKGROUND Temporary placement of self-expandable stents has been increasingly used for the management of benign esophageal diseases. OBJECTIVE To evaluate the safety of endoscopic removal of esophageal self-expandable stents placed for the treatment of benign esophageal diseases. DESIGN Multicenter retrospective study. SETTING Six tertiary care centers in the United States and Europe. ...
متن کاملPost-biliary sphincterotomy bleeding despite covered metallic stent deployment
OBJECTIVES Several endoscopic techniques have been proposed for the management of post-sphincterotomy bleeding. Lately, self-expandable metal stents deployment has gained popularity especially as a rescue therapy when other endoscopic techniques fail. METHODS-RESULTS We report the case report of a massive post-sphincterotomy bleeding in a patient with a self-expandable metal stent in the bili...