Using Hawksley random zero sphygmomanometer as a gold standard may result in misleading conclusions.
نویسندگان
چکیده
We combined a database of paired blood pressure measurements taken using the Hawksley random-zero sphygmonanometer and a standard mercury sphygmomanometer and a database of paired measurements made on a SpaceLabs 90202 ambulatory blood pressure recorder and standard sphygmomanometer to determine how the SpaceLabs 90202 would have fared if it had been assessed against the Hawksley random-zero sphygmomanometer instead of a standard sphygmomanometer. The pooled database contained 255 triplicate readings. Using the standard sphygmomanometer as gold standard, the Spacelabs had a median error of 2 mm/Hg for both systolic and diastolic. Against the Hawksley random-zero sphygmomanometer, median error was -3 mm systolic and -6 mm diastolic. The proportion of errors > 10 mm rose from 11% (systolic) and 9% (diastolic) with the standard sphygmomanometer to 16% and 29% with the Hawksley random-zero sphygmomanometer. Because it underestimates systolic and diastolic pressures, the use of the Hawksley random-zero sphygmomanometer as a gold standard may have resulted in misleading conclusions about performance of some automated BP recorders.
منابع مشابه
Field evaluation of the Copal UA-231 automatic sphygmomanometer.
Automated sphygmomanometers should be evaluated in field studies before being recommended for epidemiological use. In this study an automated sphygmomanometer, the Copal UA-231, was evaluated on the basis of duplicate blood pressure measurements, one with this machine and one with a Hawksley random zero machine, taken on 1536 participants in a health survey. The Copal-random zero differences ha...
متن کاملA comparative study of automated blood pressure recorders.
Four automated blood pressure machines (Arteriosonde 1225, Bosomat 11D, Elag Koln BE 237R and Omron HEM 3) have been compared with each other and with the Hawksley Random Zero Sphygmomanometer in 18 subjects with a wide range of BP. Each form of measurement provided a high degree of internal consistency. The Arteriosonde 1225 consistently recorded systolic and diastolic pressures which were bel...
متن کاملA comparison of the random-zero and standard mercury sphygmomanometers.
Both the standard mercury sphygmomanometer and the random-zero sphygmomanometer have been used in epidemiological studies and clinical trials. Problems arise in comparing studies since, in addition to other methodological differences, the readings obtained with the random-zero sphygmomanometer have been found to be lower than those obtained with the standard mercury sphygmomanometer. In the pre...
متن کاملProperties of the random zero sphygmomanometer.
The random zero sphygmomanometer is widely used in studies involving blood pressure measurement because it is believed to eliminate digit preference and reduce measurement error. We performed blood pressure measurements sequentially using random zero and standard sphygmomanometers in random order in 1,356 participants in the Cardiovascular Health Study. Despite adherence to the manufacturer's i...
متن کاملClinical evaluation of the Infrasonde D4000 blood pressure monitor.
Choosing a blood pressure machine from their ever increasing number is often difficult due to lack of performance data. We have evaluated a new automated non-invasive blood pressure recorder, the Infrasonde D4000. In comparison with the Hawksley random zero mercury sphygmomanometer the Infrasonde showed less variability for both systolic and diastolic readings. However, it was found to record s...