Factitiously high sodium activities on the Ektachem 400 owing to interferences by high gamma-globulin concentrations.
نویسندگان
چکیده
Sera from patients with above-normal gamma-globulin concentrations give sodium values as much as 10% higher when measured in the Kodak Ektachem 400 analyzer than results obtained with the Beckman Astra-8 analyzer. Results obtained with the Astra were in agreement with flame photometry. A similar relative bias was observed for potassium measurement in the Ektachem. This discrepancy was corrected when we used a new generation of reference fluid provided by Kodak, which is currently available for general use.
منابع مشابه
Discrepancies between sodium concentrations measured by the Kodak Ektachem 700 and by dilutional and direct ion-selective electrode analyzers.
We have identified rare (approximately 0.2% of all samples), but clinically significant, discrepancies between serum or plasma sodium concentrations measured with the Kodak Ektachem 700's direct ion-selective electrode (ISE) method and concentrations measured with two other analyzers: the Beckman Synchron CX3's dilutional ISE instrument and the Radiometer KNA2 instrument for sodium-potassium an...
متن کاملCombinedEnzymic - Jaff # { 233 } Methodfor Determinationof Creatininein Serum
Concentrations of creatinine, as determined in serum by a method involving the combined use of creatinine amidohydrolase (EC 3.5.2.10) and alkaline sodium picrate were found to be factitiously low, owing to a reversal of the enzyme reaction. This effect could be eliminated by converting creatine, the product of the enzymic reaction, to creatine phosphate. The combined enzymic-Jaff#{233} method ...
متن کاملInterferences with potentiometry of CO2 in the Ektachem 400 Analyzer.
We have investigated the Kodak Ektachem 400 Analyzer procedure for CO2 for interferences from benzyl alcohol, benzoic acid, and several compounds structurally similar to benzoic acid. Benzoic acid in plasma, at concentrations found in neonates intoxicated with benzyl alcohol, caused a large increase in the results for CO2, as did substantially above-normal concentrations of certain fatty acids ...
متن کاملTwo automated fully enzymatic assays for lipase activity in serum compared: positive interference from post-heparin lipase activity.
The fully enzymatic Wako and Ektachem assays for pancreatic lipase in serum were found to yield precise (especially Ektachem) and consistent results that were significantly correlated (r = 0.995). Less good concordance was found with the turbidimetric aca method, results being on several occasions above normal by Wako or Ektachem but within the reference limit by the aca. Supplementation of aca...
متن کاملTwo dry-reagent systems evaluated for determinations of enzyme activities.
We evaluated the Kodak Ektachem DT60/DTSC and the Boehringer Mannheim Reflotron for measuring activity concentrations of six enzymes. The Ektachem CVs for low concentrations of aspartate aminotransferase and alanine aminotransferase were high. As compared with the Ektachem and a routine "wet-chemistry" system, values for aspartate aminotransferase and alanine aminotransferase were lower as meas...