Scavenging system obstruction: A cause of raised PEEP during a magnetic resonance-guided neurosurgical procedure.
نویسندگان
چکیده
A 48-year-old man was scheduled for a temporal lobe tumor resection. The use of intraoperative magnetic resonance imaging (MRI) was required, needing special equipment including the anesthetic machine (Aestiva 5 MRI, Datex-Ohmeda, GE Healthcare, Madison, USA). After a total intravenous anesthesia induction and orotraqueal intubation the patient was ventilated with volume control modality (tidal volume 550 mL, respiratory rate 12, PEEP 6 and 1:2 inspiration/expiration ratio) ensuring normocapnia and normoxygenation controlled by arterial blood gas. Before opening the dura mater, the PEEP level suddenly rose up to 10 cm H2O (980.6 Pa), with an increase in both peak and mean airway pressures (Fig. 1, left). Simultaneously an increase of brain tension was observed in the surgical field. We ruled out a respiratory or endotracheal tube incidence because there was a lack of abnormalities in pulmonary auscultation, capnogram, pulse oximetry or hemodynamics. Our main suspect at that moment was a dysfunction in the anesthetic station. We excluded an inspiratory or expiratory valve malfunction because of the absence of rebreathing and other abnormalities in the shape of the capnography curve. Decreasing the fresh gas flow from 6 to 3 L min and changing to manual ventilation solved the air trapping. After checking all the connections between the patient and the machine and a correct aspiration from the central system, we observed a partial obstruction in one of the holes of the AGSS tube close to the anesthetic machine. The disconnection of the tube solved the problem (Fig. 1, right), confirming our hypothesis of an incident with the scavenging system. The brain tension was relieved and a successful tumor resection performed. The patient was extubated in the operation room and discharged home 5 days after the surgery without neurologic sequelae. All anesthetic machines require a security system to allow a proper scavenging system. In Aestiva 5 MRI machine the security system for avoiding an over-pressure consists on two orifices in the AGSS tube to make sure that a proper gas aspiration happens. In this case the combination of a relative high fresh gas flow and a damaged AGSS tube may have played a role in the critical incident.
منابع مشابه
بررسی مقایسه ای قدرت تشخیصی کلانژیوگرافی به روش رزونانس مغناطیسی، سونوگرافی و آزمون های آزمایشگاهی در ارزیابی های قبل از عمل جراحی انسدادهای مشکوک مجاری صفراوی
Background and purpose: Evaluation of suspected biliary obstruction is performed by common old methods such as Ultrasound, CT, and invasive cholangiography. These techniques have limitations due to the poor visualization of intraductal stones (US, CT) and the need for an invasive procedure (ERCP, PTC). Magnetic resonance cholangiography (MRC) is noninvasive imaging modality that provides good...
متن کاملProgressive visual and hearing loss as looming manifestations of spinal meningeal melanocytoma
Abstract Meningeal melanocytoma is a rare tumor of nervous system, which originates from leptomeningeal melanocytes. The locations of melanocytoma in the nervous system are most frequently in the posterior fossa or along the spinal cord, and usually appear as an extra-axial mass. The manifestations of tumor are most often due to its compressing effect on adjacent nervous structures that c...
متن کاملPerformance of magnetic field‐guided navigation system for interventional neurosurgical and cardiac procedures
A hospital-based magnetic guidance system (MGS) was installed to assist a physician in navigating catheters and guide wires during interventional cardiac and neurosurgical procedures. The objective of this study is to examine the performance of this magnetic field-guided navigation system. Our results show that the system's radiological imaging components produce images with quality similar to ...
متن کاملLarge Sphenoethmoidal Encephalocele Associated with Agenesis of Corpus Callosum and Cleft Palate
Basal encephalocele is a rare craniofacial anomaly. In the present paper we report a 10-year-old boy presented with cleft palate, congenital nystagmus, and hypertelorism. During preoperative evaluation for cleft palate repair, a pulsatile mass was detected in the pharynx. Magnetic resonance imaging showed sphenoethmoidal type of basal encephalocele and agenesis of corpus callosum. Neurosurgical...
متن کاملIntraoperative magnetic resonance imaging in neurosurgery.
Intraoperatively magnetic resonance (MR)-guided neurosurgical operations have been done since 1996, mostly for brain tumors. Several different concepts for intraoperative MRI procedures using low-, middle-, and high-field MR scanners have been reported from pioneering neurosurgical centers. In this article, we present the different solutions used in these centers from a practical point of view....
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
- Revista espanola de anestesiologia y reanimacion
دوره 62 10 شماره
صفحات -
تاریخ انتشار 2015