Impaired function of cloth-covered Starr-Edwards mitral valve prosthesis. Detection by phonocardiography.
نویسندگان
چکیده
was suspected because of phonocardiographic abnormalities. A prolonged interval from aortic closure to opening of the mitral ball suggested an impediment to movement of the ball. This was confirmed by catheterization and cinefluorography. At autopsy the impaired ball movement was found to be due to thrombus buildup on the outflow face of the mitral valve orifice. It is suggested that frequent postoperative phonocardiographic observations may be useful in detecting malfunction of valvular prostheses.
منابع مشابه
Thrombo-embolic complications of the cloth-covered Starr-Edwards prostheses No. 2300 aortic and No. 6300 mitral.
The thrombo-embolic complications of the cloth-covered Starr-Edwards prostheses No. 2300 aortic and No. 6300 mitral followed for an average of 14 months in 155 patients are reviewed. There was a high incidence of early fatal and disabling thrombo-embolus in patients 'having mitral valve replacement. Late emboli were more common after aortic valve replacement. Anticoagulant control was unsatisfa...
متن کاملRole of echocardiography and phonocardiography in diagnosis of mitral paraprosthetic regurgitation with Starr-Edwards prostheses.
Twenty-three patients who were symptomatic after mitral valve replacement, and in whom mitral paraprosthetic regurgitation was a possible cause of deterioration, were investigated by cardiac catheterization, phonocardiography, and echocardiography. Severe paraprosthetic regurgitation was present in ii patients and slight regurgitation in 2 patients. Two patients had aortic regurgitation which w...
متن کاملValve dysfunction of the cloth-covered Starr-Edwards ball valve.
BACKGROUND Between June 1968 and March 1977, Starr-Edwards cloth-covered ball valves were used for valve replacement on a routine basis. METHODS AND RESULTS Among the 66 operative survivors who underwent an isolated aortic or mitral valve replacement, 20 patients required reoperation 22 times because of valve dysfunction, thromboembolic complication, paravalvular leakage, hemolytic anemia, an...
متن کاملReoperation for a patient 25 years after a Starr-Edwards ball mitral valve was installed.
A 45-year-old female suffered from increasing dyspnea during exercise and edema of lower extremities from January 2000. She had undergone mitral valve replacement with Starr-Edwards ball prosthesis (model 6320) due to mitral valve regurgitation 25 years ago. The cardiac catheterization and echocardiography documented mitral, aortic and tricuspid valves regurgitation grade III. Left ventricular ...
متن کاملHaemolysis with Björk-Shiley and Starr-Edwards prosthetic heart valves: a comparative study.
Slater, S. D., Saflam, I. A., Bain, W. H., Turner, M. A., and Lawrie, T. D. V. (1974). Thorax, 29, 624-632. Haemolysis with Bjork-Shiley and Starr-Edwards prosthetic heart valves: a comparative study. A comparison was made of the haemolytic complications in 85 patients with two different types of Starr-Edwards cloth-covered ball and cage prosthesis with those in 44 patients with the Bjork-Shile...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
- Circulation
دوره 41 1 شماره
صفحات -
تاریخ انتشار 1970