Continuous intra-jugular venous blood-gas monitoring with the Paratrend 7 during hypothermic cardiopulmonary bypass.
نویسندگان
چکیده
We measured the accuracy of the continuous intra-vascular blood-gas monitoring system (Paratrend 7, PT7) placed in the jugular venous bulb in 18 adult patients having cardiac or aortic surgery with hypothermic cardiopulmonary bypass (CPB). After induction of anaesthesia, a PT7 sensor was inserted through a 20-gauge venous catheter into the right jugular venous bulb. Blood samples were drawn from the venous catheter and measured with a blood gas analyser (BGA). Five to eight paired measurements using the PT7 and blood samples were made per patient, and bias and precision were calculated for each patient using the Bland-Altman method. The ranges for the blood sample measurements were: pH 7.12 to 7.59, PCO(2) 3.7 to 9.6 kPa, PO(2) 3.5 to 16.0 kPa, oxygen saturation 40 to 99%, bicarbonate 18.6 to 34.4 mmol l(-1), and base excess -7.8 to 12.5 mmol l(-1). Bias and precision values were 0.014/0.071 for pH, 0/0.90 kPa for PCO(2), and -0.16/1.18 kPa for PO(2). These values were comparable with those previously made on arterial blood. However, precision for oxygen saturation in each patient varied 2.3 to 23.6% (95% CI: 6.3 to 12.9%), which was unsatisfactory for clinical measurements. Deep hypothermia ( approximately 19.6 degrees C) and marked haemodilution ( approximately 13.5%) during CPB did not influence the reliability of the PT7 sensor. Thus, we concluded that continuous intra-jugular venous blood-gas monitoring is clinically feasible using the PT7 and may provide valuable information during CPB.
منابع مشابه
Significance of continuous blood gas monitoring in cardiac surgery with cardiopulmonary bypass.
EDITOR: Standard non-invasive monitoring during anaesthesia usually includes pulse oximetry saturated pressure of oxygen (SPO2) and end-tidal carbon dioxide (end-tidal CO2). They frequently fail to monitor blood gas changes during cardiac surgery due to reduction in perfusion, ambient light, hypothermia and the use of vasoconstrictors [1]. It is also impossible to monitor SPO2 during cardiopulm...
متن کاملEffects of increasing FiO2 on venous saturation during cardiopulmonary bypass in the swine model.
Adequacy of perfusion during cardiopulmonary bypass (CPB) is dependent on nutrient delivery and waste removal from the tissue. A recent study showed that over 75% of cardiopulmonary bypass procedures are completed using continuous venous saturation (SvO2) monitoring. The purpose of this study was to determine the effect of changing FiO2concentration on SvO2. A total of eight mixed gender 45-kg ...
متن کاملDuring Hypothermic Cardiopulmonary Bypass-Correlation with Oxyhemoglobin Saturation
In 8 adult patients undergoing cardiopulmonary bypass, the perfusion flow was altered from 2.4 to 1.8 and 1.2 L/min/m during hypothermia (26.5 ± 2.3°C). Following rewarming of the patients to 37°C, the perfusion flow was also altered from 2.4 to 1.8 and 1.2 L/min/m • The resulting mixed venous oxyhemoglobin saturation (Sv0 2 ) was measured in vivo by in-line oximetry at the actual body temperat...
متن کاملContinuous fiberoptic arterial and venous blood gas monitoring in hemorrhagic shock.
STUDY OBJECTIVE To compare the performance of continuous fiberoptic blood gas monitoring with standard, intermittent blood gas sampling in the measurement of arterial and central venous blood gases during marked hemodynamic changes. DESIGN Prospective, consecutive, enrollment, experimental study. SETTING Research laboratory at a university medical center. PARTICIPANTS Seven anesthetized, ...
متن کاملEffects of Hypothermic Cardiopulmonary Bypass on Internal Jugular Bulb Venous Oxygen Saturation, Cerebral Oxygen Saturation, and Bispectral Index in Pediatric Patients Undergoing Cardiac Surgery: A Prospective Study
The objective of this study was to evaluate the effect of hypothermic cardiopulmonary bypass (CPB) on cerebral oxygen saturation (rSO2), internal jugular bulb venous oxygen saturation (SjvO2), mixed venous oxygen saturation (SvO2), and bispectral index (BIS) used to monitor cerebral oxygen balance in pediatric patients.Sixty American Society of Anesthesiologists Class II-III patients aged 1 to ...