Multicenter evaluation of Tosoh glycohemoglobin analyzer.

نویسندگان

  • I Gibb
  • A Parnham
  • M Fonfrède
  • F Lecock
چکیده

BACKGROUND We describe an Anglo-French evaluation of a new analyzer. METHODS The Tosoh HLC-723 GHb V, A1c2.2 glycohemoglobin analyzer is an HPLC instrument with primary blood tube sampling, bar-code reading, cap piercing, and the ability to chromatographically separate labile hemoglobin A1c (HbA1c). We evaluated two analytical protocols, 2.2 and 3.0 min, and compared results for blood samples collected from diabetic and nondiabetic subjects with those obtained with Bio-Rad Diamat and Variant analyzers. RESULTS Within- and between batch-imprecision (CVs) was <2% with linearity to at least 15.9% HbA1c. Although some hemoglobinopathies were detected in the 2. 2-min chromatography, clearer evidence of abnormality was visible in the 3.0-min version. Comparison with established methods showed good correlation (r = 0.993; n = 316 with Diamat; and r = 0.995; n = 133 with Variant) but highlighted calibration differences. CONCLUSIONS The problems of manual blood sample preparation, labile HbA1c, and carbamylated hemoglobin interference associated with the older instruments have been eliminated in the new Tosoh analyzer. The 3. 0-min protocol is preferred for routine use.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Screening for hemoglobinopathies during routine hemoglobin A1c testing using the Tosoh G7 Glycohemoglobin Analyzer.

Approximately 5.1% of the US population has diabetes mellitus, and hemoglobin (Hb) A1c levels are routinely measured to monitor long-term glycemic control in these patients. Many laboratories use ion exchange chromatography for such measurements, and the presence of hemoglobin variants and hemoglobinopathies often results in abnormal peaks on the chromatogram. The goal of this study was to eval...

متن کامل

Analytical verification and quality assessment of the Tosoh HLC-723GX HbA1c analyzer

OBJECTIVES Ion-exchange high-performance liquid chromatography (IE-HPLC) has long been used as a reproducible and versatile analytical tool for HbA1c measurement. In this study, we performed analytical verification and quality assessment of the recently introduced small IE-HPLC Tosoh HLC-723GX HbA1c analyzer, and a comparison of results to immunoassay (IA) and capillary electrophoresis (CE). ...

متن کامل

Incidental findings of variant hemoglobin during hemoglobin A(1c) testing.

A variant hemoglobin fraction may be an incidental finding during HbA(1c) analysis using the G8 Tosoh HPLC analyzer, but it is unclear if the retention times and fraction patterns can reliably predict the findings of a high-performance liquid chromatography (HPLC) β-thalassemia program (Bio Rad Variant II analyzer). We chose 100 samples sent for HbA(1c) determinations (G8 Tosoh) with an inciden...

متن کامل

A comparative evaluation of the analytical performances of Capillarys 2 Flex Piercing, Tosoh HLC-723 G8, Premier Hb9210, and Roche Cobas c501 Tina-quant Gen 2 analyzers for HbA1c determination

INTRODUCTION Haemoglobin A1c (HbA1c) is widely used in the management of diabetes. Therefore, the reliability and comparability among different analytical methods for its detection have become very important. MATERIALS AND METHODS A comparative evaluation of the analytical performances (precision, linearity, accuracy, method comparison, and interferences including bilirubin, triglyceride, cho...

متن کامل

Fully automated assay of glycohemoglobin with the Abbott IMx analyzer: novel approaches for separation and detection.

We describe a novel assay for measuring glycohemoglobin directly from anticoagulated whole blood with the Abbott IMx analyzer. The glycohemoglobin is labeled with a soluble polyanionic affinity reagent and the anionic complex is then captured with a cationic solid-phase matrix. Glycohemoglobin is quantified by measuring the quenching by heme of the static fluorescence from an added fluorophore....

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:
  • Clinical chemistry

دوره 45 10  شماره 

صفحات  -