Dosimetry for Small, Irregular and Rectangular Field Size for Electron Beam Therapy

نویسندگان

  • NASHAAT A. DEIAB
  • SHAYMAA Z. ABDEL KADER
  • MOHAMED EL-NAGDY
  • Nashaat A. Deiab
چکیده

Purpose: This study was to investigate the effect of irregularity in shape of field size of electron beam on the percentage depth dose and the output factor. Study the variety of the Percentage Depth Dose (PDD) and the output factors for rectangular and small field sizes when one of field dimensions less than the Practical Range (Rp) of the electron energy used. Compare the measured output factors for every rectangular and small electron lead cutout with its equivalent square field size. Material and Method: Measure the percentage depth dose PDD for small, irregular and rectangular lead cutout against reference field size (10 X 10) cm using 3D water phantom and parallel plate ion chamber on linear accelerator (Elekta precise) with electron beam energies 6, 9, 12, 15 and 18 MeV. Measure and calculate the output factor for rectangular lead cutout and its equivalent square field size. Results: The PDD parameters for irregular lead cutout varied from that for the reference field size depend on the shape of lead cutout and energy, for 6 MeV the variation ranged from 0.5 to 2%, for 9 MeV from 0.2 to 1.8%, for 12 MeV from 1.6 to 6.4%, for 15 MeV from 2.25 to 9.9% and for 18 MeV from 0.7 to 3.2%. The PDD for the small lead cutout varied from the reference field size when one of the field dimension is smaller than the practical range of the electron energy used. The variation in output for the rectangular field sizes and its equivalent square field size are ranged from 5 to 5.5% for (2 X 9) and (3 X 3), ranged from 2.3 to 4.5% for (3 X 9) and (4.5 X 4.5) and for (6 X 9) and (7 X 7) ranged from 0.1 to 0.7% at all electron energies. Conclusion: The PDD parameters and output factors for irregular, small and rectangular varied from the reference field size (10 X 10) cm2 depended on the field shape and the electron energy used.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Monte carlo simulation of varian clinac iX 10 MV photon beam for small field dosimetry

Background: The lack of lateral electronic disequilibrium (LED) becomes a main problem in small field. This factor affects the dose in target volume cannot predict correctly. In addition, utilization of high-energy linear accelerator (10 MV) can emit some unwanted particles (electron contamination). Therefore, the aim of this study was to characterize head linear accelerator (linac) Varian Clin...

متن کامل

Small photon field dosimetry using EBT2 Gafchromic film and Monte Carlo simulation

Background: Small photon fields are increasingly used in modern radiotherapy especially in intensity modulated radiation therapy (IMRT) and stereotactic radiosurgery (SRS) treatments. Accurate beam profile and central axis depth doses measurements of such beams are complicated due to the electron disequilibrium. Hence the EBT2 (external beam therapy) Gafchromic film was used for dosimetry of sm...

متن کامل

Estimation of Dosimetric Parameters based on KNR and KNCSF Correction Factors for Small Field Radiation Therapy at 6 and 18 MV Linac Energies using Monte Carlo Simulation Methods

Background: Estimating dosimetric parameters for small fields under non-reference conditions leads to significant errors if done based on conventional protocols used for large fields in reference conditions. Hence, further correction factors have been introduced to take into account the influence of spectral quality changes when various detectors are used in non-reference conditions at differen...

متن کامل

Measurement Of Scatter Factors For Small Photon Fields Using Gaf chromic EBT2 Film

Introduction: Small field dosimetry is challenging for radiotherapy dosimetry. measurement of the output factor in the air and water (Sc, Scp) is one of the input parameters for commissioning of treatment planning systems and beam modeling. The aims of this study are to measured Sc,Scp for small fields with EBT2 and Ion chamber and design a appropriate mini-phantom for small fi...

متن کامل

Physical characteristics of electron beam from conventional and beam shaper IOERT applicator: A comparison study

Introduction: Intraoperative electron radiation therapy (IOERT) is one of the cancer treatment techniques that delivers high doses to tumor bed during surgery. IOERT can be performed by either conventional LINACs or dedicated IORT accelerators such as LIAC (Light Intraoperative Accelerator). Two types of applicators can be used with LIAC dedicated accelerator including conventi...

متن کامل