Dosimetric investigation of lung tumor motion compensation with a robotic respiratory tracking system: an experimental study.

نویسندگان

  • Elena Nioutsikou
  • Yvette Seppenwoolde
  • J Richard N Symonds-Tayler
  • Ben Heijmen
  • Phil Evans
  • Steve Webb
چکیده

The benefits of using Synchrony Respiratory Tracking System (RTS) in conjunction with the CyberKnife robotic treatment device to treat a "breathing tumor" in an anthropomorphic, tissue-equivalent, thoracic phantom have been investigated. The following have been studied: (a) Synchrony's ability to allow the CyberKnife to deliver accurately a planned dose distribution to the free-breathing phantom and (b) the dosimetric implications when irregularities in the breathing cycle and phase differences between internal (tumor) and external (chest) motion exist in the course of one treatment fraction. The breathing phantom PULMONE (phantom used in lung motion experiments) has been used, which can imitate regular or irregular breathing patterns. The breathing traces from two patients with lung cancer have been selected as input. Both traces were irregular in amplitude, frequency, and base line. Patient B demonstrated a phase difference between internal and external motion, whereas patient A did not. The experiment was divided into three stages: In stage I-static, the treatment was delivered to the static phantom. In stage II-motion, the phantom was set to breathe, following the breathing trace of each of the two patients. Synchrony was switched off, so no motion compensation was made. In stage III-compensation, the phantom was set to breathe and Synchrony was switched on. A linear correspondence model was chosen to allow for phase differences between internal and external motion. Gafchromic EBT film was inserted in the phantom tumor to measure dose. To eradicate small errors in film alignment during readout, a gamma comparison with pass criteria of 3%/3 mm was selected. For a more quantitative approach, the percentage of pixels in each gamma map that exceeded the value of 1 (P1) was also used. For both breathing signals, the dose blurring caused by the respiratory motion of the tumor in stage II was degraded considerably compared with stage I (P1 = 15% for patient A and 8% for patient B). The motion compensation via the linear correspondence model was sufficient to provide a dose distribution that satisfied the set gamma criteria (P1=3% for patient A and 2% for patient B). Synchrony RTS has been found satisfactory in recovering the initial detail in dose distribution, for realistic breathing signals, even in the case where a phase delay between internal tumor motion and external chest displacement exists. For the signals applied here, a linear correspondence model provided an acceptable degree of motion compensation.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

A quantitative investigation on lung tumor site on its motion tracking in radiotherapy with external surrogates

Introduction: In external beam radiotherapy each effort is done to deliver 3D dose distribution onto the tumor volume uniformly, while minimizing the dose to healthy organs at the same time. Radiation treatment of tumors located at thorax region such as lung and liver has a challenging issue during target localization since these tumors move mainly due to respiration. There are...

متن کامل

Dynamic MLC Tracking Using 4D Lung Tumor Motion Modelling and EPID Feedback

Background: Respiratory motion causes thoracic movement and reduces targeting accuracy in radiotherapy. Objective: This study proposes an approach to generate a model to track lung tumor motion by controlling dynamic multi-leaf collimators. Material and Methods: All slices which contained tumor were contoured in the 4D-CT images for...

متن کامل

Respiratory motion effect on tumor and normal tissue doses in patients with lung cancer, treated with Intensity Modulation Radiation Therapy and Three Dimensional Conformal Radiation Therapy.

Introduction:   The aim of this study is to investigate the effect of respiratory motion during radiation therapy in patient with lung cancer and comparison of dosimetric parameters between Intensity modulation radiation therapy and three-dimensional conformal radiotherapy in lung cancer.   Materials and Methods:   Two CT scan was performred for each pati...

متن کامل

Accuracy of tumor motion compensation algorithm from a robotic respiratory tracking system: a simulation study.

The Synchrony Respiratory Tracking System (RTS) is a treatment option of the CyberKnife robotic treatment device to irradiate extra-cranial tumors that move due to respiration. Advantages of RTS are that patients can breath normally and that there is no loss of linac duty cycle such as with gated therapy. Tracking is based on a measured correspondence model (linear or polynomial) between intern...

متن کامل

Lung Tumor Tracking and Couch Control - A Feasibility Study

We look at the problem of real-time tumor tracking and motion compensation in radiotherapy techniques from a pure feedback-control point of view and try to understand it using simple mechanical models of the treatment couch and breathing patterns. Since we want to control a robotic couch based on the data collected in real-time, we propose the use of an acceleration sensor to develop a feedback...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:
  • Medical physics

دوره 35 4  شماره 

صفحات  -

تاریخ انتشار 2008