BACKGROUND Prospective studies defining the risk associated with pacemaker or implantable cardioverter-defibrillator replacement surgeries do not exist. These procedures are generally considered low risk despite results from recent retrospective series reporting higher rates. METHODS AND RESULTS We prospectively assessed predefined procedure-related complication rates associated with elective pacemaker or implantable cardioverter-defibrillator generator replacements over 6 months of follow-up. Two groups were studied: those without (cohort 1) and those with (cohort 2) a planned transvenous lead addition for replacement or upgrade to a device capable of additional therapies. Complications were adjudicated by an independent events committee. Seventy-two US academic and private practice centers participated. Major complications occurred in 4.0% (95% confidence interval, 2.9 to 5.4) of 1031 cohort 1 patients and 15.3% (95% confidence interval, 12.7 to 18.1) of 713 cohort 2 patients. In both cohorts, major complications were higher with implantable cardioverter-defibrillator compared with pacemaker generator replacements. Complications were highest in patients who had an upgrade to or a revised cardiac resynchronization therapy device (18.7%; 95% confidence interval, 15.1 to 22.6). No periprocedural deaths occurred in either cohort, although 8 later procedure-related deaths occurred in cohort 2. The 6-month infection rates were 1.4% (95% confidence interval, 0.7 to 2.3) and 1.1% (95% confidence interval, 0.5 to 2.2) for cohorts 1 and 2, respectively. CONCLUSIONS Pacemaker and implantable cardioverter-defibrillator generator replacements are associated with a notable complication risk, particularly those with lead additions. These data support careful decision making before device replacement, when managing device advisories, and when considering upgrades to more complex systems.
منابع مشابه
Arrhythmia/Electrophysiology Complication Rates Associated With Pacemaker or Implantable Cardioverter-Defibrillator Generator Replacements and Upgrade Procedures Results From the REPLACE Registry
Background—Prospective studies defining the risk associated with pacemaker or implantable cardioverter-defibrillator replacement surgeries do not exist. These procedures are generally considered low risk despite results from recent retrospective series reporting higher rates. Methods and Results—We prospectively assessed predefined procedure-related complication rates associated with elective p...
متن کاملIs correction for metallic artefacts mandatory in cardiac SPECT/CT imaging in the presence of pacemaker and implantable cardioverter defibrillator leads?
Introduction: Metallic artifacts due to pacemaker/ implantable cardioverter defibrillator (ICD) leads in CT images can produce artifactual uptake in cardiac SPECT/CT images. The aim of this study was to determine the influence of the metallic artifacts due to pacemaker and ICD leads on myocardial SPECT/CT imaging. Methods: The study included 9 patients wh...
متن کاملEffects of Upgrade Versus De Novo Cardiac Resynchronization Therapy on Clinical Response and Long-Term Survival
The beneficial impact of newly implanted cardiac resynchronization therapy (CRT) on morbidity and mortality are well described in selected patients with heart failure. Patients with heart failure already fitted with a conventional pacemaker or implantable cardioverter defibrillator (ICD) system are often considered for a CRT upgrade after the new development of CRT criteria (ie, new left bundle...
متن کاملTHE EFFECT OF NURSE-PEER-LED SUPPORT INTERVENTION ON QUALITY OF LIFE IN PATIENTS WITH AN IMPLANTABLE CARDIOVERTER-DEFIBRILLATOR: A RANDOMIZED CLINICAL TRIAL
Background & Aims: Nowadays, the quality of life and how coping and treatment of patients with an implantable cardiac defibrillator has led to concern and attempts to reduce or eliminate the problems of these patients in the people who provide health care services. This study aimed to determine the effect of nurse-peer-led support intervention on quality of life in patients with an implantable ...
متن کاملComplications associated with implantable cardioverter-defibrillator replacement in response to device advisories.
CONTEXT Recent implantable cardioverter-defibrillator (ICD) advisories and recalls have caused management dilemmas for physicians, particularly because there are no specific guidelines or data on outcomes from current management strategies. The risk of ICD generator replacement has not been assessed in this population. OBJECTIVE To determine the complication rate associated with ICD generator...