Combination balloon-ultrasound imaging catheter for percutaneous transluminal angioplasty. Validation of imaging, analysis of recoil, and identification of plaque fracture.

نویسندگان

  • J M Isner
  • K Rosenfield
  • D W Losordo
  • L Rose
  • R E Langevin
  • S Razvi
  • B D Kosowsky
چکیده

BACKGROUND We investigated the hypothesis that an ultrasound transducer positioned within an angioplasty balloon could be used to perform quantitative assessment of arterial dimensions before and after percutaneous transluminal angioplasty (PTA) and to identify certain mechanical alterations consequent to PTA, including vascular wall recoil and the initiation of plaque fractures. METHODS AND RESULTS A combination balloon-ultrasound imaging catheter (BUIC) that houses a 20-MHz ultrasound transducer within and halfway between the proximal and distal ends of an angioplasty balloon was used to perform PTA in 10 patients with peripheral vascular disease. Each PTA site was also evaluated before and after PTA by standard (nonballoon) intravascular ultrasound (IVUS) technique. In eight patients in whom satisfactory images were recorded with the BUIC before PTA, luminal cross-sectional area (XSA) of stenotic sites (0.10 +/- 0.01 cm2) did not differ significantly from measurements of XSA by IVUS (0.09 +/- 0.01 cm2, p = NS). Likewise, minimum luminal diameter (Dmin) measured by BUIC (0.34 +/- 0.02 cm) was similar to that measured by IVUS (0.33 +/- 0.01 cm, p = NS). In nine patients in whom satisfactory images were recorded with the BUIC after PTA, XSA measured by BUIC (0.29 +/- 0.03 cm2) did not differ significantly from XSA measured by IVUS (0.30 +/- 0.03 cm2, p = NS). Dmin measured by BUIC after PTA (0.57 +/- 0.02 cm) was also similar to Dmin measured by IVUS (0.57 +/- 0.03 cm, p = NS). After PTA, XSA and Dmin measured immediately after deflation were significantly less than balloon XSA and diameter at full inflation, indicating significant elastic recoil of the dilated site. For the nine patients in whom post-PTA images were satisfactory for quantitative analysis, including four patients in whom recoil was 39%, 46%, 50%, and 61%, percent recoil measured 28.6 +/- 7.2%. Finally, plaque fractures were identified on-line in six of 10 patients (60%); in each case, initiation of plaque fracture was observed at inflation pressures of 2 atm or less. CONCLUSIONS The results of this preliminary human investigation indicate that an ultrasound transducer positioned within an angioplasty balloon can be used to perform quantitative and qualitative analyses of lumen-plaque-wall alterations immediately preceding, during, and immediately after PTA in patients with peripheral vascular disease.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Catheter for Percutaneous Transluminal Angioplasty Validation of Imaging, Analysis of Recoil, and Identification of Plaque Fracture

Background. We investigated the hypothesis that an ultrasound transducer positioned within an angioplasty balloon could be used to perform quantitative assessment of arterial dimensions before and after percutaneous transluminal angioplasty (PTA) and to identify certain mechanical alterations consequent to PTA, including vascular wall recoil and the initiation of plaque fractures. Methods and R...

متن کامل

The value of myocardial perfusion imaging with Tc-99m MIBI for the prediction of perfusion improvement after percutaneous transluminal coronary angioplasty

  Introduction: Percutaneous transluminal coronary angioplasty (PTCA) is an effective method for revascularizing of stenotic coronary vessels. Lack of response to this treatment, either in symptomatic or asymptomatic patients, is usually due to incomplete revascularization, restenosis, and/or irreversibility of myocardial perfusion. Introduction of a noninvasive metho...

متن کامل

Morphological effects of coronary balloon angioplasty in vivo assessed by intravascular ultrasound imaging.

BACKGROUND Histological examination of the effects of balloon angioplasty have been described from in vitro experiments and a limited number of pathologic specimens. Intravascular ultrasound imaging permits real time cross-sectional observation of the effect of balloon dilation on the atherosclerotic plaque in vivo. METHODS AND RESULTS The morphological effects of coronary angioplasty were vi...

متن کامل

Intravascular ultrasound imaging of human coronary arteries in vivo. Analysis of tissue characterizations with comparison to in vitro histological specimens.

BACKGROUND Intravascular ultrasound imaging was performed in 27 patients after coronary balloon angioplasty to quantify the lumen and atheroma cross-sectional areas. METHODS AND RESULTS A 20-MHz ultrasound catheter was inserted through a 1.6-mm plastic introducer sheath across the dilated area to obtain real-time images at 30 times/sec. The ultrasound images distinguished the lumen from ather...

متن کامل

Validation of an automated technique for determining the mechanical characteristics of coronary arteries during balloon angioplasty: laboratory assessment with necropsy segments.

OBJECTIVES To develop a technique for automatic inflation of a percutaneous transluminal coronary angioplasty (PTCA) balloon, with continuous measurement of the balloon pressure and volume; to validate the technique for determining the mechanical characteristics of coronary arteries. METHODS During necropsy examination of the hearts of nine patients, 17 coronary artery samples were obtained f...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:
  • Circulation

دوره 84 2  شماره 

صفحات  -

تاریخ انتشار 1991