Purpose:To assess the accuracy and precision of physiological parameters estimated by the modified Tofts and Kermode (mTK) model and the adiabatic approximation to the tissue homogeneity (AATH) model in dynamic contrast enhanced (DCE) MRI

ثبت نشده
چکیده

ID: 18257 Title: Variability of Physiological Parameters Estimated by AATH and MTK Models in DCE-MRI Purpose:To assess the accuracy and precision of physiological parameters estimated by the modified Tofts and Kermode (mTK) model and the adiabatic approximation to the tissue homogeneity (AATH) model in dynamic contrast enhanced (DCE) MRI. Methods:Computer simulations were performed using the arterial input function (AIF) from Parker et al., and the multiple indicator, multiple path, indicator dilution 4 region (MMID4) model for simulating the tissue concentration time curve. A set of physiological parameters, simulating a breast tumor, was selected as the input to the MMID4. The output was then converted to signal time curve using SPGR signal equation. Four levels of Gaussian noise, corresponding to SNRs of 100, 50, 20 and 10, were added to the signal time curves, with 1000 iterations each. These curves were then transferred back to concentration time curves on which mTK and AATH models were applied for fitting. For AATH model the mean transit time (tau) was held as a fixed parameter ranged from 2 to 10 s, and the final solution was selected based on minimizing the root mean square error. Results:For common parameters obtained by both models (Ktrans , Ve, Vp), AATH provided better accuracy, which agreed with the literatures, whereas mTK resulted in superior precision. As expected, greater coefficient of variation (CV) was found with smaller SNR in all cases. For the parameters that could only be obtained with the AATH model, the CVs were less than 10% for F, E and PS, and around 20% for tau, at the SNR level of 20. Conclusions:Using the MMID4 model for data simulation, this work found that AATH model was more accurate but less precise for estimating physiological parameters as compared with mTK model. The AATH model was able to separate blood flow and permeability estimates with reasonable errors and variations.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

SU-E-I-18: Variability of Physiological Parameters Estimated by AATH and MTK Models in DCE-MRI.

PURPOSE To assess the accuracy and precision of physiological parameters estimated by the modified Tofts and Kermode (mTK) model and the adiabatic approximation to the tissue homogeneity (AATH) model in dynamic contrast enhanced (DCE) MRI. METHODS Computer simulations were performed using the arterial input function (AIF) from Parker et al., and the multiple indicator, multiple path, indicato...

متن کامل

Comparison of the Kinetic Parameters Estimated with Different Numerical Methods in Dce-mri

Introduction: Dynamic contrast enhanced MRI (DCE-MRI) has been shown its ability to noninvasively assess characterization of tumor by the kinetic parameters, such as volume transfer constant (K) and rate constant (kep). The Tofts and Kermode (TK) model [1] and modification of Tofts and Kermode (mTK) model [2] are both popular pharmacokinetic models. Nonlinear least-squares (NLSQ) method is ofte...

متن کامل

Effect of Bias in Contrast Agent Concentration Measurement on Estimated Pharmacokinetic Parameters in Brain Dynamic Contrast-Enhanced Magnetic Resonance Imaging Studies

Introduction: Pharmacokinetic (PK) modeling of dynamic contrast-enhanced magnetic resonance imaging (DCE-MRI) is widely applied in tumor diagnosis and treatment evaluation. Precision analysis of the estimated PK parameters is essential when they are used as a measure for therapy evaluation or treatment planning. In this study, the accuracy of PK parameters in brain DCE...

متن کامل

Dynamic Contrast Magnetic Resonance Imaging (DCE-MRI) and Diffusion Weighted MR Imaging (DWI) for Differentiation between Benign and Malignant Salivary Gland Tumors

Background: Salivary gland tumors form nearly 3% of head and neck tumors. Due to their large histological variety and vicinity to facial nerves, pre-operative diagnosis and differentiation of benign and malignant parotid tumors are a major challenge for radiologists. Objective: The majority of these tumors are benign; however, sometimes they tend to transform into a malignant form. Functional M...

متن کامل

Evaluation of the accuracy of dynamic contrast enhanced MRI in the diagnosis of invasive prostate neoplasm using pathological findings

Background: Prostate cancer is the most common malignancy in men and the second leading cause of death in all countries of the world. The exact mechanism of prostate cancer is not known. On the other hand, early detection of prostate cancer can lead to a complete cure. Several clinical experiments including Digital Rectum Examination (DRE), biochemistry such as Prostate Specific Antigen (PSA), ...