Biomechanical comparison of hard and soft hip protectors, and the influence of soft tissue.

نویسندگان

  • N M van Schoor
  • A J van der Veen
  • L A Schaap
  • T H Smit
  • P Lips
چکیده

INTRODUCTION Hip protectors appear to be promising in preventing hip fractures. Currently, many different hip protectors exist, and it is not clear which hip protector has the best biomechanical properties. Therefore, the objective of this study was to compare the force attenuation capacity of 10 different hip protectors. Both hard hip protectors, which primarily shunt away energy, and soft hip protectors, which primarily absorb energy, were included. METHODS Using a drop weight impact testing system and a surrogate femur, a weight of 25 kg was dropped from a height of 8 cm causing a force of almost 7,806 N on the bare femur, which simulates a severe fall. After this calibration test, soft tissue and the different hip protectors in combination with the soft tissue were tested. Each test was repeated six times. To simulate normal-weight elderly people, a 1/2-inch-thick layer of foam was chosen, reducing the force by 18%. To examine the influence of soft tissue thickness, soft tissue was also simulated by a 1-inch-thick layer of foam, reducing the force by 49%. RESULTS In the 1-inch soft tissue test, all hip protectors were capable in reducing the impact to below the average fracture threshold of elderly people (3,100 N), although the hard types performed significantly better than the soft ones (P < 0.001). In the 1/2-inch soft tissue test, only the hard hip protectors were capable of attenuating the peak force to below the average fracture threshold of 3,100 N (hard vs. soft hip protectors: P < 0.001). CONCLUSIONS This study showed that the hard, energy-shunting hip protectors were superior to the soft, energy-absorbing ones, especially in a simulation of normal-weight elderly people. With increased soft tissue thickness, soft hip protectors were also capable in reducing the impact to below the average fracture threshold of 3,100 N.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Biomechanical Comparison Between Bashti Bone Plug Technique and Biodegradable Screw for Fixation of Grafts in Ligament surgery

  Background: Ligament reconstruction is a common procedure in orthopedic surgery. Although several popular techniques are currently in use, new methods are proposed for secure fixation of the tendon graft into the bone tunnel. Purposes: We sought to introduce our new technique of Bashti bone plug for fixation of soft tissue graft in anterior cruciate ligament (ACL) reconstruction and to compar...

متن کامل

Risk of hip fractures in soft protected, hard protected, and unprotected falls.

OBJECTIVE To compare hip fracture risk in soft and hard protected falls with the risk in unprotected falls and to compare the incidence of hip fractures in nursing homes providing soft and hard hip protectors. METHODS An observational study conducted within the framework of a cluster randomized trial in 18 nursing homes. Nursing homes were randomized to offer either soft or hard hip protector...

متن کامل

Comparison of force attenuation properties of four different hip protectors under simulated falling conditions in the elderly: an in vitro biomechanical study.

The purpose of this in vitro biomechanical study is to determine the force attenuation capacity of four different hip protectors (KPH1, KPH2, Safehip, and Safetypants) in falling simulations in elderly subjects (falls to the side). The simulated falling conditions were created by a biomechanical testing system, which consisted of an impact pendulum, surrogate pelvis and femur, and two load cell...

متن کامل

Comparison of Total Fat, Percentage of Fat, Trunk Fat, Soft Lean and Mass Waist-Hip Ratio between Diabetic type II and Matched Non-Diabetic Iranians using Bioelectrical Impedance Analysis

Introduction: Research shows a relationship between diabetes (type II) and obesity. Bioelectrical Impedance Analysis (BIA) is a relatively new and approved method for the measurement of body composition. The aim of this study was to compare diabetic and non-diabetic adults on the basis of waist-hip ratio, total fat, percentage of fat, trunk fat, and soft lean mass. Methods: In this cross-sec...

متن کامل

Biomechanics of failure modalities in total hip arthroplasty

Total hip arthroplasty (THA) is the treatment of choice to relieve joint pain and loss of mobility as a result of advanced stage osteoarthritis or other hip pathologies. Despite their general success, THAs do fail, with revision rates estimated near 5% per year. Instability, defined as the complete subluxation (dislocation) of the femoral head from the acetabular socket – which usually occurs d...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:
  • Bone

دوره 39 2  شماره 

صفحات  -

تاریخ انتشار 2006