Monitoring of noninvasive ventilation by built-in software of home bilevel ventilators: a bench study.

نویسندگان

  • Olivier Contal
  • Laurence Vignaux
  • Christophe Combescure
  • Jean-Louis Pepin
  • Philippe Jolliet
  • Jean-Paul Janssens
چکیده

BACKGROUND Current bilevel positive-pressure ventilators for home noninvasive ventilation (NIV) provide physicians with software that records items important for patient monitoring, such as compliance, tidal volume (Vt), and leaks. However, to our knowledge, the validity of this information has not yet been independently assessed. METHODS Testing was done for seven home ventilators on a bench model adapted to simulate NIV and generate unintentional leaks (ie, other than of the mask exhalation valve). Five levels of leaks were simulated using a computer-driven solenoid valve (0-60 L/min) at different levels of inspiratory pressure (15 and 25 cm H(2)O) and at a fixed expiratory pressure (5 cm H(2)O), for a total of 10 conditions. Bench data were compared with results retrieved from ventilator software for leaks and Vt. RESULTS For assessing leaks, three of the devices tested were highly reliable, with a small bias (0.3-0.9 L/min), narrow limits of agreement (LA), and high correlations (R(2), 0.993-0.997) when comparing ventilator software and bench results; conversely, for four ventilators, bias ranged from -6.0 L/min to -25.9 L/min, exceeding -10 L/min for two devices, with wide LA and lower correlations (R(2), 0.70-0.98). Bias for leaks increased markedly with the importance of leaks in three devices. Vt was underestimated by all devices, and bias (range, 66-236 mL) increased with higher insufflation pressures. Only two devices had a bias < 100 mL, with all testing conditions considered. CONCLUSIONS Physicians monitoring patients who use home ventilation must be aware of differences in the estimation of leaks and Vt by ventilator software. Also, leaks are reported in different ways according to the device used.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Evaluation of ventilators for mouthpiece ventilation in neuromuscular disease.

BACKGROUND Daytime mouthpiece ventilation is a useful adjunct to nocturnal noninvasive ventilation (NIV) in patients with neuromuscular disease. The aims of the study were to analyze the practice of mouthpiece ventilation and to evaluate the performance of ventilators for mouthpiece ventilation. METHODS Practice of mouthpiece ventilation was assessed by a questionnaire, and the performance of...

متن کامل

What does built-in software of home ventilators tell us? An observational study of 150 patients on home ventilation.

BACKGROUND Recent home ventilators are equipped with built-in software which provides data such as compliance, estimations of leaks, tidal volume, minute ventilation, respiratory rate, apnea and apnea-hypopnea indexes, and percentage of inspirations triggered by the patient (or ventilator). However, for many of these variables, there is neither consensus nor documentation as to what is to be ex...

متن کامل

Effect of leak and breathing pattern on the accuracy of tidal volume estimation by commercial home ventilators: a bench study.

BACKGROUND New home ventilators are able to provide clinicians data of interest through built-in software. Monitoring of tidal volume (VT) is a key point in the assessment of the efficacy of home mechanical ventilation. OBJECTIVE To assess the reliability of the VT provided by 5 ventilators in a bench test. METHODS Five commercial ventilators from 4 different manufacturers were tested in pr...

متن کامل

Performance Characteristics of Seven Bilevel Mechanical Ventilators in Pressure-Support Mode with Different Cycling Criteria: A Comparative Bench Study

BACKGROUND Pressure support ventilation from a bilevel device is a standard technique for non-invasive home ventilation. A bench study was designed to compare the performance and patient-ventilator synchronization of 7 bilevel ventilators, in the presence of system leaks. MATERIAL/METHODS Ventilators were connected to a Hans Rudolph Series 1101 lung simulator (compliance, 50 mL/cmH2O; expirat...

متن کامل

Performance of ventilators for noninvasive positive-pressure ventilation in children.

The aim of the present study was to evaluate the performance characteristics of all the ventilators proposed for home noninvasive positive-pressure ventilation in children in France. The ventilators (one volume-targeted, 12 pressure-targeted and four dual) were evaluated on a bench which simulated six different paediatric ventilatory patterns. For each ventilator, the quality of the inspiratory...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:
  • Chest

دوره 141 2  شماره 

صفحات  -

تاریخ انتشار 2012