High-Throughput Testing of Urogenital and Extragenital Specimens for Detection of Chlamydia Trachomatis and Neisseria Gonorrhoeae with Cobas® CT/NG

نویسندگان

  • Elizabeth M. Marlowe
  • David Hardy
  • Mark Krevolin
  • Peter Gohl
  • Alexander Bertram
  • Rodney Arcenas
  • Britta Seiverth
  • Tanja Schneider
  • Oliver Liesenfeld
چکیده

We compared the analytical and clinical performance of cobas® CT/NG for use on the Cobas® 6800/8800 Systems with the Cobas® 4800 CT/NG Test from urogenital and extragenital specimens in over 12,000 specimens from both male and female subjects in Germany and the United States. The analytical sensitivity was ≤40 EB/ml for Chlamydia trachomatis (CT) and ≤1 CFU/ml for Neisseria gonorrhoeae (NG). Using clinical specimens, the overall percent agreement with the Cobas® 4800 CT/NG Test was >98.5%. Across urogenital specimens, there were 93 discrepant specimens; 76 (93.8%) of 81 CT discrepant specimens were 6800+/4800- and 10 (83.3%) of 12 NG discrepant specimens were 6800+/4800-. Sequencing verified CT results for 45 (61.6%) of 73 samples positive by 6800 and 1 (20%) of 5 positive by 4800. Similarly, 7 (70.0%) of 10 NG samples positive by 6800 and 1 of 2 positive by 4800 were confirmed by sequencing. Among discrepant extragenital specimens (all 6800+/4800-), 7 (50%) of 14 oropharyngeal and 23 (76.7%) of 30 anorectal CT discordant samples were confirmed as CT positive by sequencing; all 8 anorectal and 20 (90.9%) of 22 oropharyngeal NG discordant results were also confirmed as NG positive. In conclusion, Cobas® CT/NG for use on the Cobas® 6800/8800 Systems provides high-throughput automated solutions for sexually transmitted infection (STI) screening programs.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

The first performance report for the Bio-Rad Dx CT/NG/MG assay for simultaneous detection of Chlamydia trachomatis, Neisseria gonorrhoeae and Mycoplasma genitalium in urogenital samples.

We evaluated the clinical performance of the Bio-Rad Dx CT/NG/MG assay for the detection of Chlamydia trachomatis, Mycoplasma genitalium and Neisseria gonorrhoeae in urogenital samples in comparison with the Roche COBAS® TaqMan® CT assay for C. trachomatis and an in-house TaqMan PCR assay for M. genitalium. Swab specimens were cultured for N. gonorrhoeae. In this prospective study, urogenital s...

متن کامل

Neisseria species identification assay for the confirmation of Neisseria gonorrhoeae-positive results of the COBAS Amplicor PCR.

Screening assays for Neisseria gonorrhoeae exhibit low positive predictive values, particularly in low-prevalence populations. A new real-time PCR assay that detects and identifies individual Neisseria spp. using melt curve analysis was compared to two previously published supplementary assays. NsppID, a 16S rRNA real-time PCR/melt curve assay developed to distinguish N. gonorrhoeae from other ...

متن کامل

Detection of Chlamydia trachomatis and Neisseria gonorrhoeae in an STI population: performances of the Presto CT-NG assay, the Lightmix Kit 480 HT CT/NG and the COBAS Amplicor with urine specimens and urethral/cervicovaginal samples

OBJECTIVES This study assessed the performances of the Presto CT-NG assay, the Lightmix Kit 480 HT CT/NG and the COBAS Amplicor for Chlamydia trachomatis and Neisseria gonorrhoeae detection. DESIGN A cross-sectional study design. SETTING Izore, Centre for Diagnosing Infectious Diseases in Friesland, the Netherlands, tested samples sent from regional sexually transmitted infection (STI) outp...

متن کامل

Comparison of the cobas 4800 CT/NG test with culture for detecting Neisseria gonorrhoeae in genital and nongenital specimens in a low-prevalence population in New Zealand.

To assess the clinical utility of replacing microbial culture for Neisseria gonorrhoeae with a nucleic acid amplification test (NAAT), we compared N. gonorrhoeae culture with the cobas 4800 CT/NG test for 18,247 urogenital and 666 nongenital samples. For urogenital specimens, the sensitivity, specificity, and positive and negative predictive values of the cobas N. gonorrhoeae PCR were 98.7%, 10...

متن کامل

Validation of a laboratory-developed real-time PCR protocol for detection of Chlamydia trachomatis and Neisseria gonorrhoeae in urine.

OBJECTIVE To evaluate a sensitive and specific, real-time PCR assay with internal control for Chlamydia trachomatis and Neisseria gonorrhoeae DNA detection in urine specimens. METHODS The diagnostic performance of a laboratory-developed quadruplex assay (LDQA) targeting the cryptic plasmid and MOMP genes of C trachomatis, the porA pseudogene of N gonorrhoeae and a synthetic internal control w...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره 7  شماره 

صفحات  -

تاریخ انتشار 2017