Successful percutaneous retrieval of a leadless pacemaker due to an acute rise in pacing threshold
نویسندگان
چکیده
Leadless cardiac pacemakers (LCP) have become available recently. Both its acute and long-term performance in a large population of patients remain to be tested. Subacute rise in pacing threshold has been reported as an uncommon complication. On the other hand, the retrieval technique for LCP with passive fixation mechanism has not been previously described in details. Herein we report a newly recognized complication of an acute rise in pacing threshold very soon after implantation of an LCP without radiographic dislodgement. Percutaneous retrieval of this LCP with passive fixation mechanism was successful using a novel technique with the cryoballoon steerable sheath and a snare.
منابع مشابه
Retrieval of the Leadless Cardiac Pacemaker
Leadless cardiac pacemakers (LPs) have emerged as a safe and effective alternative to conventional transvenous pacemakers for patients with an indication for single-chamber ventricular pacing. In 2 prospective studies with intermediate follow-up, leadless pacemakers exhibited high acute implantation success rates and stable pacing performance at 6 months. Despite these promising results, the ab...
متن کاملRecent Advances in Pacing and Defibrillation
Most current pacemaker and ICD systems involve placement of lead(s) transvenously into the heart, connected to a pulse generator. Vascular access may be difficult or limited in some patients, and intravascular leads can be associated with multiple complications such as infection, lead malfunction and vascular occlusion. Recent advances in pacing and ICDs include the development of subcutaneous ...
متن کاملTreating an infected transcatheter pacemaker system via percutaneous extraction
Introduction One of the most feared complications of a traditional pacemaker system is cardiac implantable electronic device infection. In such cases, a complete removal of the entire system is recommended. With the advent of the leadless transcatheter pacemaker system (TPS), the weak link— namely, the pacing leads that contribute to this problem— can be ultimately eliminated. At least 2 major ...
متن کاملECGs and Pacemakers
Leadless pacemakers will most likely be the future of cardiac pacing. Leadless pacing employs a singlechamber system to pace only the right ventricle and is fully self-contained, which eliminates the need for leads. These systems, implanted by using a transcatheter sheath, are indicated for patients who need right ventricular pacing only. These devices enable patients to avoid lead-related comp...
متن کاملLeadless pacemaker implantation in a patient with complex congenital heart disease and limited vascular access
Management of rhythm related issues might be particularly challenging in patients with congenital heart disease due to complex anatomy and restricted vascular access. The leadless technology appears a suitable and attractive alternative for this population. We describe a patient with single ventricle physiology who successfully underwent implantation of a leadless pacemaker.
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید