Quantitative Beutler test for newborn mass screening of galactosemia using a fluorometric microplate reader.
نویسندگان
چکیده
BACKGROUND The Beutler enzyme spot test is an effective assay for newborn mass screening of galactosemia, but it is qualitative and relies on visual interpretation. We describe a quantitative, instrumental modification of the assay. METHODS We modified the macroscopic visual Beutler enzyme spot test by adding extraction of blood components from filter paper, deproteinization with acetone-methanol, and quantification and recording by a fluorescent microplate reader and personal computer. All handling was performed in microplates. The measurement time was 90 min. RESULTS Fluorescence intensity (FI) of healthy controls correlated with hematocrit and galactose-1-phosphate uridyltransferase (GALT) activity. Patients with GALT deficiency were distinguished clearly from healthy subjects and heterozygous carriers by FI. FI decreased to 75% of the initial activity after storage at 25 degrees C for 3 days and to 40% after storage at 37 degrees C for 7 days. Screening of 46 742 newborns yielded 1 false-positive result (in a heterozygous carrier), 1 patient with glucose-6-phosphate dehydrogenase deficiency, and no apparent false negatives as judged by concurrent measurements of galactose and galactose-1-phosphate. CONCLUSIONS The quantitative Beutler test can provide precise GALT activity in newborn mass screening, and can take into consideration the influence of high temperature and humidity, duration between sampling and testing, and anemia. This method is clinically useful, simple, automated, and highly reliable for newborn mass screening of galactosemia.
منابع مشابه
Microassay for screening newborns for galactosemia with use of a fluorometric microplate reader.
We describe a microassay for measuring galactose (Gal) and galactose 1-phosphate (Gal-1-P) in dried blood spots. After a coupled enzyme reaction involving galactose dehydrogenase (GADH, EC 1.1.1.48) and alkaline phosphatase (AP, EC 3.1.3.1) in a microplate well, NADH fluorescence is measured by a highly sensitive fluorometric microplate reader, capable of rapid measurement of fluorescence (2 mi...
متن کاملQuantitative Bacterial Micro-Assay for Rapid Diagnosis of Galactosemia: Application in Galactosemia Neonatal Screening
In the present study a new economic and rapid bacterial micro-assay for simultaneous detection and quantitative measurement of serum galactose was developed. Analysis of the standard curve showed a linearity range for galactose from 2 mg/dL to 180 mg/dL with a regression equation of Y = 0.013X ? 0.083; R² = 0.962. The advantage of the method is its ability to measure serum galactose quantitativ...
متن کاملCost-benefit analysis of newborn screening for galactosemia in the Philippines.
To determine the incidence of galactosemia (GAL) in the Philippines and to determine whether newborn screening for GAL is cost-beneficial from a societal perspective, cost-benefit analysis was performed. Newborn screening for GAL was done after the 24th hour of life using the Beutler test. Patients screened positive were recalled for confirmatory testing. Using incidence rates obtained from the...
متن کاملPhenylketonuria screening with a fluorometric microplate assay.
A fluorometric assay in microtitre plates for the screening of phenylketonuria was evaluated and adapted to a neonatal screening programme. Using this assay, it is possible to determine quantitatively the phenylalanine concentration in dried blood spots on filter paper. The test exhibited a linear calibration curve with a good slope as well as sufficient precision and accuracy in the statistica...
متن کاملGalactosemia diagnosis gets an upgrade.
Classic galactosemia, a defect of the metabolism of galactose, was one of the earliest defects of intermediary metabolism to be recognized in the mid 20th century (1, 2 ). The disease is caused by a defect of galactose-1phosphate uridyltransferase (GALT), an enzyme central to the Leloir pathway in which galactose is converted into glucose. Galactose is primarily derived from the lactose content...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
- Clinical chemistry
دوره 46 6 Pt 1 شماره
صفحات -
تاریخ انتشار 2000