BACKGROUND With continued increase in the use of mechanical circulatory support, the incidence of device thrombus remains a challenge. This study is a retrospective analysis of data at a single center to assess the safety and efficacy of thrombolytic use in durable mechanical assist devices. METHODS Data was analyzed retrospectively from 154 patients who underwent left ventricular assist device (LVAD) implantation from 1/1/2005 to 6/30/2014. The HMII device was implanted in 131 patients while 23 received the HVAD. LVAD thrombus was diagnosed when lactate dehydrogenase levels exceeded 1000 units/l accompanied by clinical signs of hemolysis and heart failure, echocardiographic data and surges in pump power. TPA (tissue plasminogen activator) protocol consisted of a 5 mg intravenous bolus followed by 3 mg/h infusion in normal saline for 10 h. If symptoms persisted another cycle of TPA at 1 mg/h was continued up to 48 h. RESULTS The TPA group had a 70% success rate. Success was defined as complete resolution of hemolysis and clinical symptoms with no requirement for LVAD exchange at 30 days. 95% survival was noted at 30 days and 90% were free of a hemorrhagic stroke in the TPA group. The rates of hemorrhagic strokes in the TPA group and the control group were not different (OR = 0.92). CONCLUSION The TPA protocol described here was successful consistently. Though this study is limited by its size and retrospective nature it leads the way for larger studies to generate more robust comparisons between different types of mechanical assist devices as well as the tailored use of thrombolytics in this patient population.
منابع مشابه
A phase I/II clinical trial for adult recurrent glioma using 131i-tm-601, an iodinated peptide derived from scorpion venom
131I-TM-601 is a 36-amino acid peptide, called chlorotoxin (TM-601), derived from scorpion venom labeled with I-131. TM-601 binds a receptor on the surface of tumor cells, and not on normal cells. A single dose of 131I-TM-601 administered intracranially to human xenografted mouse models of glioma has been shown to extend survival up to 269% in multiple studies. 131I-TM-601 is in a multi-center ...
متن کاملA phase I/II clinical trial for adult recurrent glioma using 131i-tm-601, an iodinated peptide derived from scorpion venom
131I-TM-601 is a 36-amino acid peptide, called chlorotoxin (TM-601), derived from scorpion venom labeled with I-131. TM-601 binds a receptor on the surface of tumor cells, and not on normal cells. A single dose of 131I-TM-601 administered intracranially to human xenografted mouse models of glioma has been shown to extend survival up to 269% in multiple studies. 131I-TM-601 is in a multi-center ...
متن کاملUtility and correlation of known anticoagulation parameters in the management of pediatric ventricular assist devices
AIM To assess utility and correlation of known anticoagulation parameters in the management of pediatric ventricular assist device (VAD). METHODS Retrospective study of pediatric patients supported with a Berlin EXCOR VAD at a single pediatric tertiary care center during a single year. RESULTS We demonstrated associations between activated thromboplastin time (aPTT) and R-thromboelastograph...
متن کامل[68Ga]Ga-PSMA-11 PET/CT for staging and patient management of high-risk prostate cancer: A single-center experience from Iran
Introduction: Accurate staging plays an important role in management of patients with prostate cancer especially in high-risk group. Today, [68Ga]Ga-PSMA-11 PET/CT should be considered as the preferred imaging tool for treatment planning and initial staging of the disease. Methods: A total number of 628 patients with prostate cancer referred to...
متن کاملThe Use of Berlin Heart EXCOR VAD in Children Less than 10 kg: A Single Center Experience
Objective: Despite the improvement in ventricular assist device (VAD) therapy in adults and in adolescents, in infant population only Berlin Heart EXCOR (BHE) is licensed as long term VAD to bridge children to Heart Transplantation (HTx). Particularly demanding in terms of morbidity and mortality are smallest patients namely the ones implanted in the first year of life or with a lower body surf...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
دوره 11 شماره
صفحات -
تاریخ انتشار 2016