Dosimetric Evaluation of Dose calculation algorithms of Monaco Treatment Planning System in the heterogeneities area

نویسندگان

  • Hassan Ali Nedaie Radiotherapy Oncology Research Centre, Cancer Institute, Tehran University of Medical Sciences, Tehran, Iran. [email protected] Department of Medical Physics and Biomedical Engineering, Tehran University of Medical Sciences, Tehran, Iran.
  • Mahbod Esfahani Imam Khomeini Hospital, Tehran University of Medical Sciences, Tehran, Iran.
  • Nasibeh Kavousi Department of Medical Physics and Medical Engineering, Faculty of Medicine, Tehran University of Medical Sciences, Tehran, Iran.
  • Somayeh Gholami Radiotherapy Oncology Research Centre, Cancer Institute, Tehran University of Medical Sciences, Tehran, Iran. [email protected]. Department of Medical Physics and Biomedical Engineering, Tehran University of Medical Sciences, Tehran, Iran. The Abdus Salam International Centre for Theoretical Physics, Strada Costiera 11, Trieste 34151, Italy.
چکیده مقاله:

Introduction: In radiation therapy, the accuracy of dose calculations by a treatment planning system (TPS) is important to achieve tumor control and to spare normal tissue. Treatment planning system calculations in the heterogeneous situation may present significant inaccuracies. In this study, three different dose calculation algorithms, pencil beam (PB), collapsed cone (CC), and Monte-Carlo (MC), provided by our planning system were compared to assess their impact on the three-dimensional planning of lung region.   Materials and Methods: The methodology was based on IAEA TEC-DOC 1583. The phantom resembling the human thorax are used and all tests were planned on three- dimensional treatment planning systems (TPSs) and irradiated with photon beams of 6, and 18 MV X-ray energies. The doses in specific points were measured with an ionization chamber. The differences between the measured and calculated doses were reported. This study was tested using different algorithms/inhomogeneity correction methods implemented in Monaco treatment planning system. Results: The measurements were conducted for all test case datasets for two photon beam energies and calculation algorithms. The deviation between the measured and calculated values for all test cases made with advanced algorithms (MC and CC) were within the agreement criteria, while the larger deviations were observed for PB algorithm. Subsequently, there were discovered dose differences greater than 20% for some simple algorithms and high energy X-ray beams. The number of measurements with results outside the agreement criteria increased with the increase of the beam energy and decreased with TPS calculation algorithm sophistication. Also, a few errors in the basic dosimetry data in TPS were detected and corrected. Conclusion: Differences were found when comparing the calculation algorithms. The PB algorithm actually overestimated the dose compared with those calculated by the CC and MC algorithms. The MC algorithm showed better accuracy than the other algorithms. Therefore advanced dose calculation algorithms are suitable and should be used in clinical practice

برای دانلود باید عضویت طلایی داشته باشید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Evaluation of Dose Calculation Accuracy of Isogray Treatment Planning System in Craniospinal Radiotherapy

Introduction: Craniospinal radiotherapy is a therapeutic technique for central nervous system (CNS) tumors, which requires meticulous attention to technique and dosimetry.Treatment planning system (TPS) is one of the main equipment in radiotherapy; therefore, the evaluation of its accuracy is essential for dose calculation. The present study evaluates the validity of Isogray TPS in craniospinal...

متن کامل

Evaluation of Dose Calculation Algorithms Accuracy for Eclipse, PCRT3D, and Monaco Treatment Planning Systems Using IAEA TPS commissioning tests in a Heterogeneous Phantom

Introduction: The accuracy of dose calculation algorithm (DCA) is highly considered in the radiotherapy sequences. This study aims at assessing the accuracy of five dose calculation algorithms in tissue inhomogeneity corrections, based on the International Atomic Energy Agency TEC-DOC 1583. Material and Methods: A heterogeneous phantom was sc...

متن کامل

Experimental evaluation of a GPU‐based Monte Carlo dose calculation algorithm in the Monaco treatment planning system

A new GPU-based Monte Carlo dose calculation algorithm (GPUMCD), devel-oped by the vendor Elekta for the Monaco treatment planning system (TPS), is capable of modeling dose for both a standard linear accelerator and an Elekta MRI linear accelerator. We have experimentally evaluated this algorithm for a standard Elekta Agility linear accelerator. A beam model was developed in the Monaco TPS (res...

متن کامل

Assessment of Dose Calculation Accuracy of TiGRT Treatment Planning System for Physical Wedged fields in Radiotherapy

Introduction Wedge modifiers are commonly applied in external beam radiotherapy to change the dose distribution corresponding to the body contour and to obtain a uniform dose distribution within the target volume. Since the radiation dose delivered to the target must be within ±5% of the prescribed dose, accurate dose calculation by a treatment planning system (TPS) is important. The objective ...

متن کامل

Evaluation of Three-dimensional Treatment Planning System (TPS) performance in dose calculation of virtual wedged fields using film dosimetry

Introduction: Nowadays radiotherapy plays an important role in cancer treatment. Different radiotherapy techniques improvement emphasizes on using of the precise ، appropriate and useful algorithms. one of these techniques are wedged which is used in radiotherapy to compensate missing tissues and create a uniform dose distribution in tissues. The Siemens Artiste linear accelera...

متن کامل

Dosimetric validation of Monaco treatment planning system on an Elekta VersaHD linear accelerator

The purpose of this study is to perform dosimetric validation of Monaco treatment planning system version 5.1. The Elekta VersaHD linear accelerator with high dose rate flattening filter-free photon modes and electron energies was used in this study. The dosimetric output of the new Agility head combined with the FFF photon modes warranted this investigation into the dosimetric accuracy prior t...

متن کامل

منابع من

با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

ذخیره در منابع من قبلا به منابع من ذحیره شده

{@ msg_add @}


عنوان ژورنال

دوره 15  شماره Special Issue-12th. Iranian Congress of Medical Physics

صفحات  54- 54

تاریخ انتشار 2018-12-01

با دنبال کردن یک ژورنال هنگامی که شماره جدید این ژورنال منتشر می شود به شما از طریق ایمیل اطلاع داده می شود.

میزبانی شده توسط پلتفرم ابری doprax.com

copyright © 2015-2023