Evaluation of Dose Calculation Accuracy of Isogray Treatment Planning System in Craniospinal Radiotherapy

نویسندگان

  • sara momeni M.S.c of Mashhad University of Medical Sciences, Medical Physics Research Center, Mashhad, Iran
  • Shokouhozaman Soleymanifard Associate Professor of Medical Physics, Research Center of Medical Physics, Mashhad University of Medical Sciences, Mashhad, Iran
چکیده مقاله:

Introduction: Craniospinal radiotherapy is a therapeutic technique for central nervous system (CNS) tumors, which requires meticulous attention to technique and dosimetry.Treatment planning system (TPS) is one of the main equipment in radiotherapy; therefore, the evaluation of its accuracy is essential for dose calculation. The present study evaluates the validity of Isogray TPS in craniospinal irradiation techniques. Material and Methods: The computed tomography (CT) images of the brain and spine of the Rando phantom were acquired. Two techniques were designed. In technique 1, the whole CNS was irradiated with 6 MV photon beam. In technique 2, the brain and spine were irradiated with 6 MV photon and 18 MeV electron beam, respectively. The tumor and organs at risk doses were measured by thermoluminescent dosimeter (TLD). In addition, photon and electron dose measurements inside and outside the treatment field were accomplished using TLD, and then compared to the corresponding values calculated by TPS. Results: According to the results, in both electron and photon beams, the differences between the doses calculated by TLD and TPS for the points inside the treatment field were less than 4% for 90% of the measurement points. However, for the points outside the treatment field borders, the differences ranged within 10-40%. These differences were indicative of the sufficient dosimetric accuracy of Isogray TPS. Conclusion: The comparison of dosimetry results with those of TPS results revealed the accuracy of Isogray TPS. In both techniques, the maximum difference between the TLD- and TPS-measured doses was observed in the mandible.

برای دانلود باید عضویت طلایی داشته باشید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Assessment of Dose Calculation Accuracy of TiGRT Treatment Planning System for Physical Wedged fields in Radiotherapy

Introduction Wedge modifiers are commonly applied in external beam radiotherapy to change the dose distribution corresponding to the body contour and to obtain a uniform dose distribution within the target volume. Since the radiation dose delivered to the target must be within ±5% of the prescribed dose, accurate dose calculation by a treatment planning system (TPS) is important. The objective ...

متن کامل

assessment of dose calculation accuracy of tigrt treatment planning system for physical wedged fields in radiotherapy

introduction wedge modifiers are commonly applied in external beam radiotherapy to change the dose distribution corresponding to the body contour and to obtain a uniform dose distribution within the target volume. since the radiation dose delivered to the target must be within ±5% of the prescribed dose, accurate dose calculation by a treatment planning system (tps) is important. the objective ...

متن کامل

Dosimetric Evaluation of Dose calculation algorithms of Monaco Treatment Planning System in the heterogeneities area

Introduction: In radiation therapy, the accuracy of dose calculations by a treatment planning system (TPS) is important to achieve tumor control and to spare normal tissue. Treatment planning system calculations in the heterogeneous situation may present significant inaccuracies. In this study, three different dose calculation algorithms, pencil beam (PB), collapsed cone (CC), ...

متن کامل

Assessment of Dose Calculation Accuracy of TiGRT Treatment Planning System for Physical Wedged fields in Radiotherapy

Introduction Wedge modifiers are commonly applied in external beam radiotherapy to change the dose distribution corresponding to the body contour and to obtain a uniform dose distribution within the target volume. Since the radiation dose delivered to the target must be within ±5% of the prescribed dose, accurate dose calculation by a treatment planning system (TPS) is important. The objective ...

متن کامل

Evaluation of Three-dimensional Treatment Planning System (TPS) performance in dose calculation of virtual wedged fields using film dosimetry

Introduction: Nowadays radiotherapy plays an important role in cancer treatment. Different radiotherapy techniques improvement emphasizes on using of the precise ، appropriate and useful algorithms. one of these techniques are wedged which is used in radiotherapy to compensate missing tissues and create a uniform dose distribution in tissues. The Siemens Artiste linear accelera...

متن کامل

منابع من

با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

ذخیره در منابع من قبلا به منابع من ذحیره شده

{@ msg_add @}


عنوان ژورنال

دوره 15  شماره 4

صفحات  231- 236

تاریخ انتشار 2018-10-01

با دنبال کردن یک ژورنال هنگامی که شماره جدید این ژورنال منتشر می شود به شما از طریق ایمیل اطلاع داده می شود.

میزبانی شده توسط پلتفرم ابری doprax.com

copyright © 2015-2023