Evaluation of Effect of Different Computed Tomography Scanning Protocols on Hounsfield Unit and Its Impact on Dose Calculation by Treatment Planning System

نویسندگان

  • Anshu Singh Department of Radiotherapy, Delhi State Cancer Institute(s), Dilshad Garden, Delhi – 110095, India
  • Mamta Mahur Department of Radiotherapy, Delhi State Cancer Institute(s), Dilshad Garden, Delhi – 110095, India
  • Munendra Singh Department of Physics, School of Basic Sciences & Research, Sharda University, Greater Noida, Uttar Pradesh – 201306, India
  • Om Prakash Gurjar Roentgen-SAIMS Radiation Oncology Centre, Sri Aurobindo Institute of Medical Sciences
  • PS Negi Department of Radiotherapy, Delhi State Cancer Institute(s), Dilshad Garden, Delhi – 110095, India
  • Richa Sharma Department of Radiotherapy, Delhi State Cancer Institute(s), Dilshad Garden, Delhi – 110095, India
  • RK GROVER Department of Radiotherapy, Delhi State Cancer Institute(s), Dilshad Garden, Delhi – 110095, India
چکیده مقاله:

Introduction: In radiotherapy treatment planning system (TPS), basic input is the data from computed tomography (CT) scan, which takes into account the effect of inhomogeneities in dose calculations. Measurement of CT numbers may be affected by scanner-specific parameters. Therefore, it is important to verify the effect of different CT scanning protocols on Hounsfield unit (HU) and its impact on dose calculation. This study was carried out to analyse the effect of different tube voltages on HU for various tissue substitutes in phantom and their dosimetric impact on dose calculation in TPS due to variation in HU–relative electron density (RED) calibration curves. Materials and Methods: HU for different density materials was obtained from CT images of the phantom acquired at various tube voltages. HU-RED calibration curves were drawn from CT images with various tissue substitutes acquired at different tube voltages used to quantify the error in dose calculation for different algorithms. Doses were calculated on CT images acquired at 120 kVp and by applying CT number to RED curve obtained from 80, 100, 120, and 140 kVp voltages. Results: No significant variation was observed in HU of different density materials for various kVp values. Doses calculated with applying different HU-RED calibration curves were well within 1%. Conclusion: Variation in doses calculated by algorithms with various HU-RED calibration curves was found to be well within 1%. Therefore, it can be concluded that clinical practice of using the standard HU-RED calibration curve by a 120 kVp CT acquisition technique is viable.

برای دانلود باید عضویت طلایی داشته باشید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

an investigation of the impact of self monitoring on langauge teachers motivational practice and its effect on learners motivation

the central purpose of this study was to conduct a case study about the role of self monitoring in teacher’s use of motivational strategies. furthermore it focused on how these strategies affected students’ motivational behavior. although many studies have been done to investigate teachers’ motivational strategies use (cheng & d?rnyei, 2007; d?rnyei & csizer, 1998; green, 2001, guilloteaux & d?...

the effect of functional/notional approach on the proficiency level of efl learners and its evaluation through functional test

in fact, this study focused on the following questions: 1. is there any difference between the effect of functional/notional approach and the structural approaches to language teaching on the proficiency test of efl learners? 2. can a rather innovative language test referred to as "functional test" ge devised so so to measure the proficiency test of efl learners, and thus be as much reliable an...

15 صفحه اول

the geopolitics of afghanistan after september 11, 2001, and its impact on iranian national security

abstract u.s. attack to afghanistan after 9/11 provided opportunities for us who tried to become the dominant power after collapse of ussr and bipolar world. thinkers like fukuyama believed american liberal democrats thought would be universally accepted without resistances. liberal democracy is the best and final model of governance; in this regard, united state has moral responsibilities for...

15 صفحه اول

منابع من

با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

ذخیره در منابع من قبلا به منابع من ذحیره شده

{@ msg_add @}


عنوان ژورنال

دوره 14  شماره 3

صفحات  149- 154

تاریخ انتشار 2017-09-01

با دنبال کردن یک ژورنال هنگامی که شماره جدید این ژورنال منتشر می شود به شما از طریق ایمیل اطلاع داده می شود.

میزبانی شده توسط پلتفرم ابری doprax.com

copyright © 2015-2023