Relationship between the optimum cut off frequency for Butterworth filter and lung-heart ratio in 99mTc myocardial SPECT

نویسندگان

چکیده مقاله:

Background: We investigated whether the lungheart ratio parameter (LHR) can be used to identify the optimum cut off frequency for Butterworth filter in 99mTc myocardial SPECT imaging. Materials and Methods: This study involved a cardiac phantom system consisting of cardiac insert in which 1.10 cm cold defect was inserted into its myocardium wall and filled with 4.0 μCi/ml (0.148 MBq/ml) 99mTc concentration. The cardiac insert was then put into a cylindrical tank which filled with six different 99mTc concentrations as background. Thus, six target background concentrations ratios (T/B) were carried out. The LHR was determined for every SPECT raw image obtained corresponding to each T/B. Then, 130 different combinations of filter parameters from Butterworth filter were utilized to reconstruct each SPECT raw image. The determination of count in myocardium, background, and defect regions of interest (ROI) were performed for every reconstructed image. All the count values were then used to calculate contrast, signal-to-noise ratio (SNR), and defect size. Each criterion was graded (1 to 100) and then summed together to obtain total grade. The optimum cut off frequency for each LHR was determined from the total grade. The relation between optimum cut off frequency for Butterworth filter and LHR was established using linear regression. Results: There were good relationship between the optimum Butterworth cut off frequency and LHR (R2 = 0.864, p < 0.01). The optimal cut off frequency correspond to the change in LHR can be expressed by the equation: Optimum cut off frequency = 0.715*LHR + 0.227. Conclusion: This study suggests that the optimum cut off frequency for Butterworth filter should be determined by referring to LHR in each patient study. Iran. J. Radiat. Res., 2010 8 (1): 17-24

برای دانلود باید عضویت طلایی داشته باشید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

relationship between the optimum cut off frequency for butterworth filter and lung-heart ratio in 99mtc myocardial spect

background: we investigated whether the lungheart ratio parameter (lhr) can be used to identify the optimum cut off frequency for butterworth filter in 99mtc myocardial spect imaging. materials and methods: this study involved a cardiac phantom system consisting of cardiac insert in which 1.10 cm cold defect was inserted into its myocardium wall and filled with 4.0 μci/ml (0.148 mbq/ml) 99mtc c...

متن کامل

Determination of the optimum filter for qualitative and quantitative 99mTc myocardial SPECT imaging

Background: Butterworth, Gaussian, Hamming, Hanning, and Parzen are commonly used SPECT filters during filtered back-projection (FBP) reconstruction, which greatly affect the quality and size accuracy of image. Materials and Methods: This study involved a cardiac phantom in which 1.10 cm thick cold defect was inserted into its myocardium wall and filled with 4.0 μCi/ml (0.148 MBq/ml) 99mTc conc...

متن کامل

Determination of optimum filter in myocardial SPECT: A phantom study

Background: In myocardial perfusion SPECT imaging, images are degraded by photon attenuation, the distance-dependent collimator, detector response and photons scatter. Filters greatly affect quality of nuclear medicine images Materials and Methods: A phantom simulating heart left ventricle was built. About 1mCi of 99mTc was injected into the phantom. Images were taken from this phantom. Some fi...

متن کامل

Filter selection for 99mTc-Sestamibi myocardial perfusion SPECT imaging [Persian]

Myocardial perfusion SPECT imaging with Tc99m-sestamibi is the most accurate non-invasive means of detecting coronary artery disease and assessing the severity of perfusion abnormalities in the patients with coronary stenosis. Though simple and straight forward the results produced by the technique is very much affected by the details being used. One of the main problems with the test, wh...

متن کامل

determination of the optimum filter for qualitative and quantitative 99mtc myocardial spect imaging

background: butterworth, gaussian, hamming, hanning, and parzen are commonly used spect filters during filtered back-projection (fbp) reconstruction, which greatly affect the quality and size accuracy of image. materials and methods: this study involved a cardiac phantom in which 1.10 cm thick cold defect was inserted into its myocardium wall and filled with 4.0 μci/ml (0.148 mbq/ml) 99mtc conc...

متن کامل

determination of optimum filter in myocardial spect: a phantom study

background: in myocardial perfusion spect imaging, images are degraded by photon attenuation, the distance-dependent collimator, detector response and photons scatter. filters greatly affect quality of nuclear medicine images materials and methods: a phantom simulating heart left ventricle was built. about 1mci of 99mtc was injected into the phantom. images were taken from this phantom. some fi...

متن کامل

منابع من

با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

ذخیره در منابع من قبلا به منابع من ذحیره شده

{@ msg_add @}


عنوان ژورنال

دوره 8  شماره None

صفحات  17- 24

تاریخ انتشار 2010-06

با دنبال کردن یک ژورنال هنگامی که شماره جدید این ژورنال منتشر می شود به شما از طریق ایمیل اطلاع داده می شود.

کلمات کلیدی

میزبانی شده توسط پلتفرم ابری doprax.com

copyright © 2015-2023