Highly Selective Perchlorate Membrane Electrode Based on Synthesized Schiff Base Complex
Authors
Abstract:
In this work we introduce a practical use of synthesized Schiff base metal complex in sensing application. A new perchlorate-selective electrode based on the complex N, N'-4-nitro-phenylmethanebis-(salicylaldiminato) copper (II) [CuL] as the membrane carrier was developed. The electrode exhibits a good Nernstian slope of -55.0 ± 1.0 mV/decade and a linear range of for perchlorate. The limit of detection is . It has a fast response time <15s and can be used for more than three months. The selective coefficients of Perchlorate-selective electrode were determined by the Fixed Interference Method (FIM) and Fixed Primary ion Method (FPM), and could be used in the pH range 4.0-9.0. It was employed as an indicator electrode for direct determination of perchlorate in real samples.
similar resources
My Resources
Journal title
volume 29 issue 3
pages 123- 132
publication date 2010-09-01
By following a journal you will be notified via email when a new issue of this journal is published.